Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to
Yanjun Hua1, Wei Qiu1, Qiuyi Xiao1
1Department of Ophthalmology, Shanghai Jiao Tong University Affiliated Sixth People's Hospital, Xuhui District, Shanghai, China.
Purpose:
To assess the precision (repeatability and reproducibility) of ocular parameters measured by the Tomey OA-2000 biometer, and to compare them with those measured by the IOLMaster.
Methods:
In this prospective study, the right eyes of 108 healthy subjects were included. Three consecutive scans were obtained by 2 observers using the Tomey OA-2000, and in the same session one observer used the IOLMaster (version 5.4.4.0006) for the measurements. About 1 week later, 3 scans were obtained by one observer using the Tomey OA-2000. The axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), keratometer readings, pupil diameter (PD) and corneal diameter (CD) values measured by the Tomey OA-2000 and IOLMaster were analyzed. The coefficient of variation (CoV), intraclass correlation coefficient (ICC), within subject standard deviation (Sw) and 2.77Sw were calculated to assess the repeatability and reproducibility. The paired t test and Bland-Altman plots were used to analyze the differences and agreements of parameters measured by the two devices, respectively.
Results:
Intraobserver repeatability, and interobserver and intersession reproducibility of the AL, CCT, ACD, LT, Kf, Ks, Km, PD and CD values measured by the Tomey OA-2000 biometer showed a CoV of less than 1% except that for PD, and an ICC of more than 0.97 except that for PD and CD. The AL, Kf, Ks, Km and CD values measured by the Tomey OA-2000 were 0.058 ± 0.094 mm, 0.088± 0.150 diopters (D), 0.163 ± 0.170 D, 0.127 ± 0.117 D and 0.171 ± 0.217 mm lower than those measured by the IOLMaster, respectively (all Ps < 0.05). However, the ACD values from the two devices were comparable (P = 0.169). The 95% linite of agreement (LoA) of the AL, ACD, CD and all keratometer readings were no more than 0.24 mm, 0.14 mm 0.60 mm and 0.5 D, respectively.
Conclusion:
Except for the PD and CD, the ocular parameters measured by the Tomey OA-2000 were highly repeatable and reproducible. Except for the CD value, there was good agreement of ocular parameters measured by the Tomey OA-2000 and the IOLMaster in healthy eyes.
More Related Videos
05:29A Highly Reproducible and Straightforward Method to Perform In Vivo Ocular Enucleation in the Mouse after Eye Opening
Published on: October 6, 2014
07:26Characterizing the Relationship Between Eye Movement Parameters and Cognitive Functions in Non-demented Parkinson's Disease Patients with Eye Tracking
Published on: September 26, 2019
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Wave Parameters
Comparing Copy Number Variations and SNPs
Copy number variations or CNVs are the structural variations that cover more than 1kb of DNA sequence. The single nucleotide polymorphism (SNP), on the other hand, is a single nucleotide change or a point mutation that is found in more than 1%...
Accuracy and Precision
The Precise Definition of a Limit
Comparative Excretory Systems