Patient radiation dose and fluoroscopy time during ERCP: a single-center, retrospective study of influencing factors
Ekaterina Saukko1, Juha M Grönroos2,3, Paulina Salminen2,3
1a The Medical Imaging Centre of Southwest Finland, Turku University Hospital , Turku , Finland.
Objectives:
Recently, both the number and the complexity with associated increased technical difficulty of therapeutic ERCP procedures have significantly increased resulting in longer procedural and fluoroscopy times. During ERCP, the patient is exposed to ionizing radiation and the consequent radiation dose depends on multiple factors. The aim of this study was to identify factors affecting fluoroscopy time and radiation dose in patients undergoing ERCP.
Materials And Methods:
Data related to patient demographics, procedural characteristics and radiation exposure in ERCP procedures (n = 638) performed between August 2013 and August 2015 was retrospectively reviewed and analyzed. Statistically significant factors identified by univariate analyses were included in multivariate analysis with fluoroscopy time (FT) and dose area product (DAP) as dependent variables. Effective dose (ED) was estimated from DAP measurements using conversion coefficient.
Results:
The factors independently associated with increased DAP during ERCP were age, gender, radiographer, complexity level of ERCP, cannulation difficulty grade, bile duct injury and biliary stent placement. In multivariate analysis the endoscopist, the complexity level of ERCP, cannulation difficulty grade, pancreatic duct leakage, bile duct dilatation and brushing were identified as predictors for a longer FT. The mean DAP, FT, number of acquired images and ED for all ERCP procedures were 2.33 Gy·cm2, 1.84 min, 3 and 0.61 mSv, respectively.
Conclusions:
Multiple factors had an effect on DAP and FT in ERCP. The awareness of these factors may help to predict possible prolonged procedures causing a higher radiation dose to the patient and thus facilitate the use of appropriate precautions.
Related Concept Videos
Patient-centered Care
Other Factors Affecting Respiration Centers
However, the ability to hold one's breath voluntarily is not limitless. When the CO2 concentration in the blood reaches a critical...
Bioavailability Study Design: Single Versus Multiple Dose Studies
Biological Effects of Radiation
Bioavailability: Influencing Factors
Chemical Factors Affecting Respiration Centers
CO2 has a potent influence on respiration and is strictly regulated....