Quantification of Iodine Concentration Using Single-Source Dual-Energy Computed Tomography in a Calf Liver
Objective:
To evaluate the accuracy of single-source dual-energy computed tomography (ssDECT) in iodine quantification using various segmentation methods in an ex vivo model.
Methods:
Ten sausages, injected with variable quantities of iodinated contrast, were inserted into 2 livers and scanned with ssDECT. Material density iodine images were reconstructed. Three radiologists segmented each sausage. Iodine concentration, volume, and absolute quantity were measured. Agreement between the measured and injected iodine was assessed with the concordance correlation coefficient (CCC). Intrareader agreement was assessed using the intraclass correlation coefficient (ICC).
Results:
Air bubbles were observed in sausage (IX). Sausage (X) was within the same view as hyper-attenuating markers used for localization. With IX and X excluded, CCC and ICC were greater than 0.98 and greater than 0.88. When included, CCC and ICC were greater than 0.94 and greater than 0.79.
Conclusions:
Iodine quantification was reproducible and precise. However, accuracy reduced in sausages consisting of air filled cavities and within the same view as hyperattenuating markers.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
08:41Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Free Energy
Activation Energy
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Expressing Solution Concentration
Concentrations may be quantitatively assessed using a wide variety of measurement units, each convenient for particular applications. Molarity (M) is a useful concentration unit for many applications in chemistry.