Imaging of symptomatic total knee arthroplasty with cone beam computed tomography
Antti Jaroma1, Juha-Sampo Suomalainen2, Lea Niemitukia2
11 Department of Orthopaedics, Traumatology and Hand Surgery, Kuopio University Hospital, Kuopio, Finland.
Background:
Computed tomography (CT) can be used to assess the rotational alignment of prosthesis components to identify possible underlying causes of symptomatic total knee arthroplasty (TKA). The use of cone beam computed tomography (CBCT) for the imaging of extremities is relatively new, although it has been widely used in dental imaging.
Purpose:
To assess the intra- and inter-observer reproducibility of CBCT, as well as to validate CBCT for TKA component and periprosthetic bone diagnostics.
Material And Methods:
CBCT scans were performed on 18 patients the day before a scheduled revision TKA, from which the intra- and inter-observer reproducibility were assessed. Component rotation and loosening were evaluated. Perioperative bone defects were classified.
Results:
The inter-observer intraclass coefficient correlation (ICC) for femoral component rotation was 0.41 (95% confidence interval [CI] = 0.12-0.69). For the tibial component, the ICC was 0.87 (95% CI = 0.74-0.94). Intra-observer reproducibilities were 0.70 (95% CI = 0.35-0.87) and 0.92 (95% CI = 0.80-0.97), respectively. The sensitivity for tibial component loosening was 97% and the specificity was 85%. The reliability of bone defect classification was only weak to moderate.
Conclusion:
Two-dimensional (2D) CBCT scanning provides reliable and reproducible data for determining the rotation of femoral and tibial components, while showing minor overestimation of tibial component loosening. CBCT is a promising new tool for the evaluation of symptomatic knee arthroplasty patients, with a substantially lower radiation dose compared to conventional 2D multi-slice CT.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Deflection of a Beam
Singularity functions, described in an earlier lesson, are powerful mathematical tools that represent discontinuities within a function commonly encountered in structural loading...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET