The Perinatal Asphyxiated Lamb Model: A Model for Newborn Resuscitation
Published on: August 15, 2018
Delivery of positive end-expiratory pressure to preterm lambs using common resuscitation devices
Marta Thio1,2,3,4, Jennifer A Dawson1,2,3, Kelly J Crossley5,6
1Newborn Research Centre, The Royal Women's Hospital, Melbourne, Victoria, Australia.
Background:
In neonatal resuscitation, a ventilation device providing positive end-expiratory pressure (PEEP) is recommended. There is limited information about PEEP delivery in vivo, using different models of self-inflating bag (SIB) at different inflation rates and PEEP settings.
Methods:
We compared PEEP delivery to intubated preterm lambs using four commonly available models of paired SIBs and PEEP valves, with a T-piece, with gas flow of 8 L/min. Peak inspiratory pressure inflations of 30 cmH2O, combined with set PEEP of 5, 7 and 10 cmH2O, were delivered at rates of 20, 40 and 60/min. These combinations were repeated without gas flow. We measured mean PEEP, maximum and minimum PEEP, and its difference (PEEP reduction).
Results:
A total of 3288 inflations were analysed. The mean PEEP delivered by all SIBs was lower than set PEEP (P<0.001), although some differences were <0.5 cmH2O. In 55% of combinations, the presence of gas flow resulted in increased PEEP delivery (range difference 0.3-2 cmH2O). The mean PEEP was closer to set PEEP with faster inflation rates and higher set PEEPs. The mean (SD) PEEP reduction was 3.9 (1.6), 8.2 (1.8), 2 (0.6) and 1.1 (0.6) cmH2O with the four SIBs, whereas it was 0.5 (0.2) cmH2O with the T-piece.
Conclusions:
PEEP delivery with SIBs depends on the set PEEP, inflation rate, device model and gas flow. At recommended inflation rates of 60/min, some devices can deliver PEEP close to the set level, although the reduction in PEEP makes some SIBs potentially less effective for lung recruitment than a T-piece.
Related Concept Videos
Common Ion Effect
Cardiopulmonary Resuscitation I: Adult
Design Example: Identifying the Locations of Monuments in the Field Using Global Positioning System Device
Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices
Cardiopulmonary Resuscitation III: AED Use
Cardiopulmonary Resuscitation IV: Pharmacological Management