Comparison of visual outcomes and flap morphology using 2 femtosecond-laser platforms
Montserrat Garcia-Gonzalez1, Carmen Bouza-Miguens1, Alberto Parafita-Fernandez1
1From the Clínica Novovisión Madrid (Garcia-Gonzalez, Gros-Otero, Villa-Collar, Teus), the Clínica Rementería (Garcia-Gonzalez), Universidad Europea de Madrid (Bouza-Miguens, Villa-Collar), Hospital Universitario Príncipe de Asturias (Gros-Otero, Cañones-Zafra, Teus), University of Alcalá, Alcalá de Henares, Madrid, and Hospital Provincial de Pontevedra (Parafita-Fernandez), Pontevedra, Spain.
Purpose:
To compare the 3-month postoperative refractive results and flap thickness homogeneity after correction of myopia using the Intralase femtosecond platform and the Victus dual femtosecond laser platform.
Setting:
Novovision Clinic, Madrid, Spain.
Design:
Prospective case series.
Methods:
Eyes treated with a 60 kHz femtosecond laser (Group 1) and eyes treated with a dual femtosecond laser (Group 2) were included. The refractive results 3 months postoperatively were compared. The differences between the maximum and the minimum thickness point of each flap, the intraflap standard deviation (SD), and the achieved and the targeted central flap thickness were evaluated in each group.
Results:
The study comprised 31 eyes in Group 1 and 20 eyes in Group 2. Three months postoperatively, the uncorrected distance visual acuity, residual refraction, efficacy, safety, and predictability were significantly better in Group 1. The dual femtosecond laser induced a significantly greater increase in higher-order aberrations than the conventional femtosecond laser. The mean central flap thickness was 123.2 μm ± 9.2 (SD) in Group 1 and 116.3 ± 14.4 μm in Group 2 (P = .005). The mean difference between the maximum and the minimum flap thickness points in each flap was significantly lower in Group 1 than in Group 2 (22 ± 8.3 μm versus 35.7 ± 16.9 μm) (P = .0007). The intraflap SD was significantly lower in Group 1 than in Group 2 (6.1 μm versus 10.8 μm) (P = .0001).
Conclusion:
The conventional femtosecond laser provided better efficacy, safety, predictability, and flap thickness homogeneity than the dual femtosecond laser for the correction of myopia.
Related Concept Videos
Predicting Reaction Outcomes
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Outcomes of Glycolysis
Cellular respiration can occur aerobically (with oxygen) or anaerobically (without oxygen). In the presence of oxygen, cellular respiration starts with glycolysis and continues with pyruvate...
Guidelines for Writing Outcome
Patient outcomes reflect the patient's response to the goal rather than what the nurse aims to achieve. Terminology should be observable and measurable to avoid the reader's interpretation. The desired outcome should be realistic and achievable in the designated care timeframe. Expected outcomes should align with adjunctive therapies. The outcome should enhance care...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Microbial Morphologies