Volumetric Measurements of Brain Shift Using Intraoperative Cone-Beam Computed Tomography: Preliminary Study
Vitor Mendes Pereira1,2, Iris Smit-Ockeloen3, Olivier Brina1
1Division of Neuroradiology, Department of Medical Imaging, University Hospitals of Geneva, Geneva, Switzerland.
Background:
Cerebrospinal fluid leakage and ventricular compression during open surgery may lead to brain deformation called brain shift. Brain shift may affect intraoperative navigation that is based on image-based preoperative planning. Tools to correct or predict these anatomic modifications can be important to maintain precision during open guided neurosurgery.
Objective:
To obtain a reliable intraoperative volumetric deformation vector field describing brain shift during intracranial neurosurgical procedures.
Methods:
We acquired preoperative and intraoperative cone-beam computed tomography enhanced with intravenous injection of iodine contrast. These data sets were preprocessed and elastically registered to obtain the volumetric brain shift deformation vector fields.
Results:
We obtained the brain shift deformation vector field in 9 cases. The deformation fields proved to be highly nonlinear, particularly around the ventricles. Interpatient variability was considerable, with a maximum deformation ranging from 8.1 to 26.6 mm and a standard deviation ranging from 0.9 to 4.9 mm.
Conclusion:
Contrast-enhanced cone-beam computed tomography provides a feasible technique for intraoperatively determining brain shift deformation vector fields. This technique can be used perioperatively to adjust preoperative planning and coregistration during neurosurgical procedures.
More Related Videos
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Deflection of a Beam
Singularity functions, described in an earlier lesson, are powerful mathematical tools that represent discontinuities within a function commonly encountered in structural loading...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Prismatic Beams: Problem Solving
The design begins with analyzing the beam as a free body to identify moments and force balances, thereby determining support reactions. Next, the...