Reduction of respiratory ghosting motion artifacts in conventional two-dimensional multi-slice Cartesian turbo
Yuuji Inoue1,2, Masami Yoneyama3, Masanobu Nakamura4
1Tsukuba Imaging Center, Ibaraki, Japan.
Abstract:
The two-dimensional Cartesian turbo spin-echo (TSE) sequence is widely used in routine clinical studies, but it is sensitive to respiratory motion. We investigated the k-space orders in Cartesian TSE that can effectively reduce motion artifacts. The purpose of this study was to demonstrate the relationship between k-space order and degree of motion artifacts using a moving phantom. We compared the degree of motion artifacts between linear and asymmetric k-space orders. The actual spacing of ghost artifacts in the asymmetric order was doubled compared with that in the linear order in the free-breathing situation. The asymmetric order clearly showed less sensitivity to incomplete breath-hold at the latter half of the imaging period. Because of the actual number of partitions of the k-space and the temporal filling order, the asymmetric k-space order of Cartesian TSE was superior to the linear k-space order for reduction of ghosting motion artifacts.
More Related Videos
Related Concept Videos
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Cartesian Vector Notation
NMR Spectroscopy: Spin–Spin Coupling
Spin–Spin Coupling: One-Bond Coupling
Cartesian Form for Vector Formulation
Vector Components in the Cartesian Coordinate System