Consistency of Multiple Renal Functional MRI Measurements Over 18 Months
Lu-Ping Li1,2, Jon Thacker3, Wei Li1,2
1Radiology, Northshore University HealthSystem, Evanston, Illinois, USA.
Background:
Identification of patients with progressive chronic kidney disease (CKD) and those likely to respond to candidate therapeutics is urgently needed. Functional MRI measurements have shown promise. However, knowledge about the consistency of the measurements is essential to conduct longitudinal studies.
Purpose/Hypothesis:
To investigate the consistency of repeated functional MRI measurements in healthy subjects.
Study Type:
Prospective, longitudinal study.
Subjects:
Seventeen healthy subjects were examined on two different occasions, 18 months apart.
Field Strength/Sequence:
Multiple gradient-recalled-echo, 2D navigator-gated flow-sensitive alternating inversion recovery True-FISP and spin-echo planar diffusion-weighted sequences were used on a 3T scanner. Images were acquired on two different scanner configurations.
Assessment:
Blood oxygenation level-dependent (BOLD) R2*, arterial spin labeling (ASL) perfusion-derived blood flow (BF) and apparent diffusion coefficient (ADC) maps were analyzed using a custom image processing toolbox. Regions of interest (ROIs) were placed on renal cortex, medulla, and whole kidney. Multiple researchers were involved in defining the ROIs.
Statistical Tests:
Intra- and intersubject coefficients of variation (CV) and Bland-Altman plots were used to measure consistency and evaluate bias in the measurements. A nonparametric Wilcoxon test was used to compare differences between two timepoints.
Results:
The intrasubject CV for R2* and ADC were 6.8% and 5.3% with small (-3.8 and 5.3%) bias, respectively, comparing baseline and 18-month data. Intrasubject CV for renal cortex BF was higher (18.7%) compared to R2* and ADC, but comparable to prior literature values over shorter durations. It also exhibited a larger bias (-15.4%) between two timepoints and significantly lower values (P = 0.022) at 18-month data.
Data Conclusion:
All three MRI parameters over 18 months, even with a scanner upgrade and involving multiple observers, showed good consistency. These results are useful for the interpretation of longitudinal data and support the use of these methods to monitor progression in patients with CKD.
Level Of Evidence:
1 Technical Efficacy: Stage 1 J. MAGN. RESON. IMAGING 2018;48:514-521.
Related Concept Videos
Serum Studies: Renal Function Tests
Drug Dosing in Renal Diseases: Measurement of Glomerular Filtration Rate
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...
Multiple Allele Traits
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Renal Corpuscle
Glomerulus: Structure and Function
The glomerulus is a tiny, intricate network of capillaries located at the beginning of the nephron. It's enveloped by the Bowman's capsule and receives its blood supply from an afferent arteriole, which divides into numerous...