External implantable defibrillator as a bridge to reimplant after explant for infection: Experience from two centers
Gabriele Dell'Era1, Matteo Ziacchi2, Eleonora Prenna1
1Division of Cardiology, University of Eastern Piedmont, Maggiore della Carità Hospital, Novara, Italy.
Background:
The management of patients explanted for implantable converter defibrillator (ICD) infections may be complex when anti-bradycardia pacing and tachyarrhythmia protection are needed. We aimed to test the efficacy and safety of a conventional ICD externally connected to a transvenous dual-coil lead as bridging therapy before the reimplantation.
Methods And Results:
We enrolled seven patients explanted for ICD infections and needed prolonged antibiotic therapy in two high-volume hospitals in Italy and treated them with a passive-can external ICD for a mean of 13 (4-30) days before reimplant. One patient experienced an electrical storm, efficaciously recognized by the external ICD and treated with antitachycardia pacing and shocks. On-demand pacing was granted for all the patients. No device-related complications were reported.
Conclusions:
An external ICD seems safe and efficacious as a bridge to reimplant in patients explanted for ICD infections.
Related Concept Videos
Design Example: Strain Gauge Bridge or Wheatstone Bridge
Wheatstone Bridge
Thus, for accurate resistance measurements, a...
Bridge rectifier
Operationally, the bridge rectifier allows current flow through two of its diodes during each...
Cross-bridge Cycle
Lattice Centering and Coordination Number
Types of Unit Cells
Imagine taking a large number of identical...
Work Done on a System by External Force
In the presence of a non-conservative opposing force, like friction, some part of the work done...