Cavopulmonary mechanical circulatory support in Fontan patients and the need for physiologic control: A computational
Marcus Granegger1, Martin Schweiger1, Marianne Schmid Daners2
11 Pediatric Heart Center, University Children's Hospital, University of Zurich, Zurich, Switzerland.
Purpose:
Rotary blood pumps are a promising treatment approach for patients with a total cavopulmonary connection and a failing cardiovascular system. The aim of this study was to investigate the hemodynamic effects of cavopulmonary support using a numerical model with closed-loop baroreflex and exercise mechanisms.
Methods:
A numerical model of the univentricular cardiovascular system was developed, mimicking the hemodynamics during rest and exercise. Rotary blood pumps with different hydraulic pump characteristics (flat vs steep pressure-flow relationships) were investigated in the cavopulmonary position. Furthermore, two support modes-a constant speed setting and a physiologically controlled speed-were examined.
Results:
Hemodynamics without rotary blood pumps were achieved with less than 10% deviation from reported values during rest and exercise. Rotary blood pumps at constant speed improve the hemodynamics at rest, however, they constitute a hydraulic resistance during light (steep characteristics) or moderate (flat characteristics) exercise. In contrast, physiologic control increases cardiac output (moderate exercise: 8.2 vs 7.4 L/min) and reduces sympathetic activation (heart rate at moderate exercise: 111 vs 123 bpm).
Conclusion:
In this simulation study, the necessity of an automatically controlled rotary blood pump in the cavopulmonary position was shown. A pump at constant speed might constitute an additional resistance to venous return during physical activity. Therefore, a physiologic control algorithm based on the pressure difference between the caval veins and the atrial pressure is proposed to improve hemodynamics, especially during physical activity.
Related Concept Videos
Open and closed-loop control systems
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal...
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Anatomy of the Circulatory System
Physiology of Respiration II: Neurogenic Control of Respiration
Central Control
The brainstem is the primary site of central control, hosting respiratory centers:
Clearance Models: Physiological Models
The organ's clearance rate depends on the blood flow to the organ and the extraction ratio (E). The extraction ratio describes the organ's...
Imaging Studies III: Computed Tomography