Serial intravascular ultrasound evaluation of the DESolve™ novolimus-eluting bioresorbable coronary scaffold system
Gentil Barreira1, J Ribamar Costa1, Ricardo Costa1
1Department of Invasive Cardiology, Instituto Dante Pazzanese de Cardiologia, São Paulo, Brazil.
Introduction:
Bioresorbable coronary scaffolds (BRS) have been developed to increase the late safety of coronary angioplasty by providing transitory coronary support and then being fully incorporated to the vessel wall. In the present trial, we sought to evaluate the performance and changes over time in the DESolve™ novolimus-eluting BRS using serial intravascular ultrasound (IVUS) in patients submitted to percutaneous coronary interventions.
Methods:
Single-center, prospective, non-randomized study involving 17 consecutive patients submitted to implantation of the DESolve™ scaffold and serial evaluated with IVUS at different time points (post procedure, 6 and 18 months). Primary endpoint included the variation in lumen, scaffold, and vessel diameter and area along the months.
Results:
Angiographic success was achieved in all cases. Scaffold area and volume on IVUS were significantly greater at 6 months than at baseline (6.41 ± 1.35 mm2 vs. 7.35 ± 1.53 mm2 , P < 0.002; and 101.19 ± 20.9 mm3 vs. 118.51 ± 26.6 mm3 , P = 0.001). Late lumen loss was 0.22 ± 0.30 mm at 6 months and 0.33 ± 0.44 mm at 18 months. No major adverse cardiac events occurred. At late follow-up (18 months) the scaffold was not visualized by IVUS; therefore the evaluation was restricted to lumen and vessel at that time point.
Conclusions:
IVUS serial evaluation of the novel DESolve BRS showed an increase in the device dimensions between baseline and 6 months, with concomitant enlargement of lumen dimensions and effective suppression of neointimal proliferation. At 18 months, no footprint of the scaffold was detected and the initial lumen enlargement was sustained.
Related Concept Videos
Serial Position Effect
High-Performance Liquid Chromatography: Elution Process
Coronary Circulation
Coronary circulation begins at the base of the aorta, where two main arteries arise—the left and right coronary arteries. These arteries encircle the heart in the coronary sulcus and supply the...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Coronary Artery Disease I: Introduction
Coronary Artery Disease II: Pathophysiology