A critical evaluation of the Beckman Coulter Access hsTnI: Analytical performance, reference interval and concordance
Carel J Pretorius1, Jillian R Tate2, Urs Wilgen1
1Department of Chemical Pathology, Pathology Queensland, Brisbane, Australia; University of Queensland, Brisbane, Australia.
Introduction:
We investigated the analytical performance, outlier rate, carryover and reference interval of the Beckman Coulter Access hsTnI in detail and compared it with historical and other commercial assays.
Materials And Methods:
We compared the imprecision, detection capability, analytical sensitivity, outlier rate and carryover against two previous Access AccuTnI assay versions. We established the reference interval with stored samples from a previous study and compared the concordances and variances with the Access AccuTnI+3 as well as with two commercial assays.
Results:
The Access hsTnI had excellent analytical sensitivity with the calibration slope 5.6 times steeper than the Access AccuTnI+3. The detection capability was markedly improved with the SD of the blank 0.18-0.20 ng/L, LoB 0.29-0.33 ng/L and LoD 0.58-0.69 ng/L. All the reference interval samples had a result above the LoB value. At a mean concentration of 2.83 ng/L the SD was 0.28 ng/L (CV 9.8%). Carryover (0.005%) and outlier (0.046%) rates were similar to the Access AccuTnI+3. The combined male and female 99th percentile reference interval was 18.2 ng/L (90% CI 13.2-21.1 ng/L). Concordance amongst the assays was poor with only 16.7%, 19.6% and 15.2% of samples identified by all 4 assays as above the 99th, 97.5th and 95th percentiles. Analytical imprecision was a minor contributor to the observed variances between assays.
Conclusion:
The Beckman Coulter Access hsTnI assay has excellent analytical sensitivity and precision characteristics close to zero. This allows cTnI measurement in all healthy individuals and the capability to identify numerically small differences between serial samples as statistically significant. Concordance in healthy individuals remains poor amongst assays.
Related Concept Videos
Critical Numbers and the Closed Interval Method
Kendall's Coefficient of Concordance
Prediction Intervals
However, the point estimate is most likely not the exact value of the population parameter, but close to it. After calculating point estimates, we construct interval estimates, called confidence intervals or prediction intervals. This prediction interval comprises a range of values unlike the point estimate and is a better predictor of the observed sample value, y.