Bioimpedance analysis is safe in patients with implanted cardiac electronic devices

Xavier Chabin1, Ouarda Taghli-Lamallem1, Aurélien Mulliez2

  • 1Université Clermont Auvergne, Cardio Vascular Interventional Therapy and Imaging (CaVITI), Image Science for Interventional Techniques (ISIT), UMR6284, and CHU Clermont-Ferrand, Cardiology Department, F-63003 Clermont-Ferrand, France.

Insights

Bioimpedance analysis (BIA) is safe for patients with cardiac implantable electronic devices (CIEDs). This study found no electromagnetic interference, suggesting BIA can be used to monitor conditions like heart failure in these individuals.

Area of Science:

  • Biomedical Engineering
  • Cardiology
  • Medical Devices

Background:

  • Increasing prevalence of cardiac implantable electronic devices (CIEDs) globally.
  • Current guidelines contraindicate bioimpedance analysis (BIA) due to potential electromagnetic interference (EMI) risks.
  • BIA is valuable for monitoring heart failure and nutritional disorders, but CIEDs limit its use.

Purpose of the Study:

  • To evaluate the safety and potential for EMI of BIA in patients with CIEDs.
  • To determine if BIA measurements affect CIED function.
  • To assess the feasibility of using BIA in CIED patients for clinical management.

Main Methods:

  • 200 patients with CIEDs were enrolled in the study.
  • BIA was performed during routine CIED interrogations.
  • Key CIED parameters (battery voltage, lead impedance, pacing thresholds, electrograms) were monitored before and after BIA.

Main Results:

  • No significant changes in CIED battery voltage, lead impedance, or pacing thresholds were observed during BIA.
  • No instances of oversensing or undersensing were detected in intracardiac electrograms.
  • Follow-up assessments at 6 and 12 months showed no adverse effects on CIEDs.

Conclusions:

  • Bioimpedance analysis (BIA) does not cause electromagnetic interference in patients with cardiac implantable electronic devices (CIEDs).
  • BIA can be safely performed in patients with CIEDs, expanding its clinical utility.
  • This finding supports revising guidelines to include BIA for managing conditions in CIED patients.
Abstract

Related Concept Videos

Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices01:28

Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices

Parenteral drug delivery systems play a crucial role in modern therapeutics by enabling the direct administration of drugs into the systemic circulation, bypassing the gastrointestinal tract. These systems are particularly valuable for poorly absorbed oral medications that are unstable in the digestive environment or require rapid onset or sustained therapeutic levels. Delivery is achieved through intravenous, intramuscular, or subcutaneous routes, each selected based on the drug's properties...
4
Guidelines and Strategies for Safe Computer Charting01:18

Guidelines and Strategies for Safe Computer Charting

The guidelines and strategies provided by the American Nurses Association (ANA) and the Canadian Nurses Association (CNA) offer essential principles for ensuring safe and secure computer charting systems in healthcare settings. Let's break down each recommendation:
Maintain Confidentiality and Security:
2.7K
Electron Carriers01:24

Electron Carriers

Electron carriers can be thought of as electron shuttles. These compounds can easily accept electrons (i.e., be reduced) or lose them (i.e., be oxidized). They play an essential role in energy production because cellular respiration is contingent on the flow of electrons.
Over the many stages of cellular respiration, glucose breaks down into carbon dioxide and water. Electron carriers pick up electrons lost by glucose in these reactions, temporarily storing and releasing them into the electron...
92.1K
Electron Affinity03:07

Electron Affinity

The electron affinity (EA) is the energy change for adding an electron to a gaseous atom to form an anion (negative ion).
43.7K
Electron Behavior00:54

Electron Behavior

Overview
Electrons are negatively charged subatomic particles that are attracted to an orbit around the positively-charged nucleus of an atom. They reside in locations that are associated with energy levels called shells and are further organized into sub-shells and orbitals within each shell.
Electrons Orbit the Nucleus
Electrons are found in specific locations outside of the nucleus. The shell in which an electron resides indicates the general energy level of the electron: those closer to the...
109.8K
Electron Transport Chains01:28

Electron Transport Chains

The final stage of cellular respiration is oxidative phosphorylation that consists of two steps: the electron transport chain and chemiosmosis. The electron transport chain is a set of proteins found in the inner mitochondrial membrane in eukaryotic cells. Its primary function is to establish a proton gradient that can be used during chemiosmosis to produce ATP and generate electron carriers, such as NAD+ and FAD, that are used in glycolysis and the citric acid cycle.
The ETC is comprised of...
113.1K