Measuring Blood Pressure in Mice using Volume Pressure Recording, a Tail-cuff Method
Published on: May 15, 2009
Reliability of single office blood pressure measurements
Thilo Burkard1,2, Michael Mayr1, Clemens Winterhalder3
1Hypertension Clinic, ESH Hypertension Centre of Excellence, Medical Outpatient Department, University Hospital Basel, Basel, Switzerland.
Objectives:
Standard operating procedures for office blood pressure measurement (OBPM) vary greatly between guidelines and studies. We aimed to compare the difference between a single OBPM and the mean of the three following measurements. Further, we studied how many patients with possible hypertension may be missed due to short-term masked hypertension (STMH) and how many might be overdiagnosed due to short-term white coat hypertension (STWCH).
Design And Setting:
In this cross-sectional, single-centre trial, 1000 adult subjects were enrolled. After 5 min of rest, four sequential standard OBPMs were performed at 2 min intervals in a quiet room in sitting position. We compared the first (fBPM) to the mean of the second to fourth measurement (mBPM). STMH was defined as fBPM <140 mm Hg systolic and <90 mm Hg diastolic and mBPM systolic ≥140 mm Hg or diastolic ≥90 mm Hg. STWCH was defined as fBPM systolic ≥140 mm Hg or diastolic ≥90 mm Hg and mBPM <140 mm Hg systolic and <90 mm Hg diastolic.
Results:
Complete measurements were available in 802 subjects. Between fBPM and mBPM, 662 (82.5%), 441 (55%) and 208 (25.9%) subjects showed a difference in systolic and 531 (66.2%), 247 (30.8%) and 51 (6.4%) in diastolic blood pressure (BP) values of >2 mm Hg, >5 mm Hg and >10 mm Hg, respectively. In 3.4% of initially normotensives STMH and in 34.3% of initially hypertensives, STWCH was apparent.
Conclusions:
There are significant differences between a single OBPM and the mean of consecutive BP measurements. Our study provides evidence that a single OBPM should not be the preferred method and should be discouraged in future guidelines.
Trial Registration Number:
NCT02552030;Results.
Related Concept Videos
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Blood Pressure
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the...
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer