On the accuracy of cochlear duct length measurement in computed tomographic images
G Jakob Lexow1, Marcel Kluge2, Nils-Claudius Gellrich3
1Department of Otolaryngology and Cluster of Excellence "Hearing4all" EXC 1077/1, Hannover Medical School, Carl-Neuberg-Str. 1, 30625, Hanover, Germany. lexow.jakob@mh-hannover.de.
Purpose:
Patient specific selection of cochlear implants would benefit from pre-operative knowledge of cochlear length. Several methods for its measurement or estimation have been described in literature. This study focused on the achievable accuracy in clinically available imaging.
Methods:
Five simplified cochlea models milled into porcine bone were scanned in water using clinical cone beam computed tomography. Due to their well-known dimensions these phantoms served as gold standard for the length measurements. Each phantom was measured ten times using the custom software Comet. In addition, cochleae in ten image datasets taken indiscriminately from clinical routine were measured ten times each to test the precision under realistic conditions. The results were also compared to estimations based on the diameter of the basal turn (A value) as described in literature.
Results:
Measurement accuracy of the phantoms' lengths was high (average error: - 0.2 mm; standard deviation: 0.3 mm). The pooled standard deviation for the measurements in clinical datasets was 0.6 mm. Errors resulted mainly from problems locating the helicotrema. The estimations differed on average - 1.7 to + 0.4 mm from the manual measurements and had standard deviations between 0.5 and 0.6 mm depending on the algorithm.
Conclusions:
The program Comet was successfully used to accurately measure the length of the cochlea models in clinically available imaging. The lower image quality of patient scans reduced the precision of the measurement. Estimations using the A value are a quicker alternative for averagely sized cochleae in cases where the lack of accuracy is tolerable.
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Improving Translational Accuracy
Improving Translational Accuracy
Imaging Studies III: Computed Tomography
Accuracy and Precision
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...