Assessing matrix, interferences and comparability between the Abbott Diagnostics and the Beckman Coulter
Peter A Kavsak1,2, Paul Malinowski3, Chantele Roy3
1McMaster University, Juravinski Hospital and Cancer Centre, 711 Concession Street Hamilton, L8V 1C3 ON, Canada.
Background:
Analytical evaluation of high-sensitivity cardiac troponin (hs-cTn) assays, with particular attention to imprecision, interferences and matrix effects, at normal cTn concentrations, is of utmost importance as many different clinical algorithms use concentration cutoffs <10 ng/L for decision-making. The objective for the present analytical study was to compare the new Beckman Coulter hs-cTnI assay (Access hsTnI) to Abbott's hs-cTnI assay in different matrices and for different interferences, with a focus on concentrations <10 ng/L.
Methods:
The limit of blank (LoB) and the limit of detection (LoD) were determined in different matrices for the Beckman hs-cTnI assay. Passing-Bablok regression and difference plots were determined for 200 matched lithium heparin and EDTA plasma samples for the Beckman assay and 200 lithium heparin samples for the Abbott assay. Both EDTA and heparin plasma samples were also evaluated for stability under refrigerated conditions, for endogenous alkaline phosphatase interference and for hemolysis and icterus.
Results:
The Beckman hs-cTnI assay LoB was 0.5 ng/L with the following range of LoDs=0.8-1.2 ng/L, with EDTA plasma yielding lower concentrations as compared to lithium heparin plasma (mean difference=-14.9%; 95% CI=-16.9 to 12.9). Below 10 ng/L, lithium heparin cTnI results from the Beckman assay were on average 1.1 ng/L (95% CI=0.7 to 1.5) higher than the Abbott results, with no difference between the methods when using EDTA plasma (mean difference =-0.1 ng/L; 95% CI=-0.3 to 0.2). Low cTnI concentrations were less effected by interferences in EDTA plasma.
Conclusions:
The Access hsTnI method can reliably detect normal cTnI concentrations with both lithium heparin and EDTA plasma being suitable matrices.
More Related Videos
10:12A Sensitive and Specific Quantitation Method for Determination of Serum Cardiac Myosin Binding Protein-C by Electrochemiluminescence Immunoassay
Published on: August 8, 2013
07:40Impact of High-intensity Interval Exercise and Moderate-Intensity Continuous Exercise on the Cardiac Troponin T Level at an Early Stage of Training
Published on: October 10, 2019
Related Concept Videos
Interference and Diffraction
RNA Interference
This process occurs naturally in cells, often through the activity of genomically-encoded microRNAs. Researchers can take advantage of this mechanism by introducing synthetic RNAs to deactivate specific genes for research or therapeutic purposes. For example, RNAi could be used...
The Extracellular Matrix
Interference and Decay
Interference occurs when competing memories hinder the retrieval of particular information. It can be classified into two types: proactive and retroactive interference. Proactive...
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Sound Waves: Interference