Day-and-Night Closed-Loop Insulin Delivery in a Broad Population of Pregnant Women With Type 1 Diabetes: A Randomized
Zoe A Stewart1,2, Malgorzata E Wilinska1, Sara Hartnell2
1Wellcome Trust-Medical Research Council Institute of Metabolic Science, University of Cambridge, Cambridge, U.K.
Objective:
Despite advances in technology, optimal glucose control remains elusive and neonatal complications remain ubiquitous in type 1 diabetes (T1D) pregnancy. Our aim was to examine the safety, efficacy, and longer-term feasibility of day-and-night closed-loop insulin delivery.
Research Design And Methods:
We recruited 16 pregnant women (mean [SD]: age 32.8 [5.0] years, T1D duration 19.4 [10.2] years, HbA1c 8.0% [1.1], and BMI 26.6 [4.4] kg/m2) to an open-label, randomized, crossover trial. Participants completed 28 days of closed-loop and sensor-augmented pump (SAP) insulin delivery separated by a washout period. Afterward, participants could continue to use the closed-loop system up to 6 weeks postpartum. The primary end point was the proportion of time with glucose levels within the target range (63-140 mg/dL).
Results:
The proportion of time with glucose levels within target was comparable during closed-loop and SAP insulin delivery (62.3 vs. 60.1% [95% CI -4.1 to 8.3]; P = 0.47). Mean glucose and time spent hyperglycemic >140 mg/dL also did not differ (131.4 vs. 131.4 mg/dL [P = 0.85] and 36.6 vs. 36.1% [P = 0.86], respectively). During closed-loop, fewer hypoglycemic episodes occurred (median 8 [range 1-17] vs. 12.5 [1-53] over 28 days; P = 0.04) and less time at <63 mg/dL (1.6 vs. 2.7%; P = 0.02). Hypoglycemia <50 mg/dL (0.24 vs. 0.47%; P = 0.03) and low blood glucose index (1.0 vs. 1.4; P = 0.01) were lower. Less nocturnal hypoglycemia (2300-0700 h) during closed-loop therapy (1.1 vs. 2.7%; P = 0.008) and a trend toward higher overnight time in target (67.7 vs. 60.6%; P = 0.06) were found.
Conclusions:
Closed-loop insulin delivery was associated with comparable glucose control and significantly less hypoglycemia than SAP therapy. Larger, longer-duration multicenter trials are now indicated to determine clinical efficacy of closed-loop insulin delivery in T1D pregnancy and the impact on neonatal outcomes.
Related Concept Videos
Open and closed-loop control systems
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal...
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into...
Diabetes Mellitus: Type 2 and Gestational
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Types of Hormones
Crossover Experiments
Crossover designs are performed even with smaller sample sizes since the samples can act as their controls. These are better than simple randomized trials since patients are exposed to all the treatments.