Related Experiment Video
Updated: Feb 13, 2026

A Protocol for Detecting and Scavenging Gas-phase Free Radicals in Mainstream Cigarette Smoke
Published on: January 2, 2012
Optimized method for determination of 16 FDA polycyclic aromatic hydrocarbons (PAHs) in mainstream cigarette smoke by
Jana Jeffery1, Maria Carradus2, Karolina Songin2
1British American Tobacco, Research and Development, Southampton, UK. jana_jeffery@bat.com.
Abstract:
A gas chromatography-mass spectrometry (GC-MS) method was validated for the determination of 16 polycyclic aromatic hydrocarbons (PAHs) from the FDA list of 93 harmful or potentially harmful constituents of mainstream cigarette smoke (MCS). Target analytes were extracted from total particulate matter using accelerated solvent extraction with a toluene/ethanol solvent mixture. Matrix artefacts were removed by two-step solid-phase extraction process. Three different GC-MS systems [GC-MS (single quadrupole), GC-MS/MS (triple quadrupole) and GC-HRMS (high resolution, magnetic sector)] using the same separation conditions were compared for the analysis of MCS of 3R4F Kentucky reference cigarettes generated under ISO and intense smoking regimes. The high mass resolution (m/∆m ≥ 10,000) and associated selectivity of detection by GC-HRMS provided the highest quality data for the target PAHs in MCS. Owing to the HR data acquisition mode enabling measurement of accurate mass, limits of quantification for PAHs were 5 to 15-fold lower for GC-HRMS than for GC-MS/MS and GC-MS. The presented study illustrates that the optimised sample preparation strategy followed by GC-HRMS analysis provides a fit-for-purpose and robust analytical approach allowing measurement of PAHs at (ultra)low concentrations in MCS. Furthermore, the study illustrates the importance and benefits of robust sample preparation and clean-up to compensate for limited selectivity when low-resolution MS is used.
More Related Videos
04:39A Modified QuEChERS-HPLC Method for Detection of Polycyclic Aromatic Hydrocarbons in Zebrafish Embryos Exposed to Fine Particulate Matter
Published on: June 13, 2025
09:30A Microcontroller Operated Device for the Generation of Liquid Extracts from Conventional Cigarette Smoke and Electronic Cigarette Aerosol
Published on: January 18, 2018
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Mass Spectrometry: Aromatic Compound Fragmentation
Aromatic Hydrocarbon Anions: Structural Overview
Due to the absence of continuous...
Aromatic Hydrocarbon Cations: Structural Overview
Removing one hydrogen from the intervening CH2 group...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Chemical Stoichiometry and Gases: Using Ideal Gas Law to Determine Moles