An ex vivo study of automated motion artefact correction and the impact on cone beam CT image quality and
Rubens Spin-Neto1, Louise H Matzen1, Lars W Schropp1
11 Oral Radiology, Department of Dentistry and Oral Health, Aarhus University , Aarhus , Denmark.
Objectives:
To assess the impact of head motion artefacts and an automated artefact-correction system on cone beam CT (CBCT) image quality and interpretability for simulated diagnostic tasks.
Methods:
A partially dentate human skull was mounted on a robot simulating four types of head movement (anteroposterior translation, nodding, lateral rotation, and tremor), at three distances (0.75, 1.5, and 3 mm) based on two movement patterns (skull returning/not returning to the initial position). Two diagnostic tasks were simulated: dental implant planning and detection of a periapical lesion. Three CBCT units were used to examine the skull during the movements and no-motion (control): Cranex 3Dx (CRA), Orthophos SL 3D (ORT), and X1 without (X1wo) and with (X1wi) an automated motion artefact-correction system. For each diagnostic task, 88 examinations were performed. Three observers, blinded to unit and movement, scored image quality: presence of stripe artefacts (present/absent), overall unsharpness (present/absent), and image interpretability (interpretable/non-interpretable). κ statistics assessed interobserver agreement, and descriptive statistics summarized the findings.
Results:
Interobserver agreement for image interpretability was good (average κ = 0.68). Regarding dental implant planning, X1wi images were interpretable by all observers, while for the other units mainly the cases with tremor were non-interpretable. Regarding detection of a periapical lesion, besides tremor, most of the 3 mm movements based on the "not returning" pattern were also non-interpretable for CRA, ORT, and X1wo. For X1wi, two observers scored 1.5 mm tremor and one observer scored 3 mm tremor as non-interpretable.
Conclusions:
The automated motion artefact-correction system significantly enhanced CBCT image quality and interpretability.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Impact Loading on a Cantilever Beam
When an object is dropped onto the free end of a cantilever, its potential energy due to gravity is...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Distance Corrections
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...