Using Laser Doppler Imaging and Monitoring to Analyze Spinal Cord Microcirculation in Rat
Published on: May 30, 2018
Laser doppler imaging as additional monitoring after digital replanting: A prospective study
Marc Schmid1, S Morteza Seyed Jafari2,3, Luzian Haug1
1Department for Plastic, Reconstructive and Hand Surgery, Inselspital, Bern University Hospital, Bern, Switzerland.
Background:
Despite various exisiting monitoring methods, there is still a need for new technologies to improve the quality of post-operative evaluation of digital replantation. The purpose of the study is using a laser Doppler imaging device (Easy-LDI) as an additional tool to assess perfusion. In this method, the changes in the frequency of the laser ligth provide information regarding perfusion of the monitored tissue.
Patients And Methods:
This study included seven patients (10 fingers; age of patients: 21-57 years) who suffered from a total (n = 6) or subtotal amputation (n = 4) due to accidents. In addition to hourly standard monitoring with clinical evaluation and skin thermometry, revascularized fingers were hourly monitored with Easy LDI for 48 h.
Results:
LDI measurement values ranged between 0.8 and 223 (mean 90.62 ± 21.42) arbitrary perfusion units (APU). The mean LDI values before and after revascularization were 7.1 ± 2.85 and 65.30 ± 30.83 APU, respectively. For the successful revascularized fingers (8 of 10 fingers) values from 19 to 223 APU (mean 98.52 ± 15.48) were demonstrated. All of the replants survived, but due to venous occlusion two digits required revision 12 and 35 h after revascularization, respectively. In the two cases, Easy-LDI also showed a constant and slow decline of the perfusion values. Furthermore, Pearson normalized correlation coefficient showed a positive significant correlation between temperatures of the replants and LDI-values (P < .001, r = +0.392) and a negative significant correlation between Δtemperature and LDI-values (P < .001, r = -0.474).
Conclusion:
The LDI-device might be a promising additional monitoring technique in detection of perfusion disturbance in monitoring digital replantations.
Related Concept Videos
Doppler Effect - I
Doppler Effect - II
Conjugate Addition (1,4-Addition) vs Direct Addition (1,2-Addition)
Conjugate addition results in a thermodynamically stable product. The reaction retains the stronger C=O bond at the expense of the weaker C=C π bond. The process is slow as the β carbon is less electrophilic than the carbonyl carbon.
Direct addition products are...
Conjugate Addition of Enolates: Michael Addition
Additional Subnuclear Structures
The nucleus contains many membrane-less subnuclear organelles or nuclear bodies, such as nucleoli, Cajal bodies, speckles,...
Additional Subnuclear Structures