Application of Ultrasound and Shear Wave Elastography Imaging in a Rat Model of NAFLD/NASH
Published on: April 20, 2021
Clinical acceptance testing and scanner comparison of ultrasound shear wave elastography
Zaiyang Long1, Donald J Tradup1, Pengfei Song1
1Department of Radiology, Mayo Clinic, Rochester, MN, USA.
Abstract:
Because of the rapidly growing use of ultrasound shear wave elastography (SWE) in clinical practices, there is a significant need for development of clinical physics performance assessment methods for this technology. This study aims to report two clinical medical physicists' tasks: (a) acceptance testing (AT) of SWE function on ten commercial ultrasound systems for clinical liver application and (b) comparison of SWE measurements of targets across vendors for clinical musculoskeletal application. For AT, ten GE LOGIQ E9 XDclear 2.0 scanners with ten C1-6-D and ten 9L-D transducers were studied using two commercial homogenous phantoms. Five measurements were acquired at two depths for each scanner/transducer pair by two operators. Additional tests were performed to access effects of different coupling media, phantom locations and operators. System deviations were less than 5% of group mean or three times standard deviation; therefore, all systems passed AT. A test protocol was provided based on results that no statistically significant difference was observed between using ultrasound gel and salt water for coupling, among different phantom locations, and that interoperator and intraoperator coefficient of variation was less than 3%. For SWE target measurements, two systems were compared - a Supersonic Aixplorer scanner with a SL10-2 and a SL15-4 transducer, and an abovementioned GE scanner with 9L-D transducer. Two stepped cylinders with diameters of 4.05-10.40 mm were measured both longitudinally and transaxially. Target shear wave speed quantification was performed using an in-house MATLAB program. Using the target shear wave speed deduced from phantom specs as a reference, SL15-4 performed the best at the measured depth. However, it was challenging to reliably measure a 4.05 mm target for either system. The reported test methods and results could provide important information when dealing with SWE-related tasks in the clinical environment.
More Related Videos
12:18Magnetic Resonance Elastography Methodology for the Evaluation of Tissue Engineered Construct Growth
Published on: February 9, 2012
05:56Implementation of Non-invasive Point of Care Transient Elastography for Evaluation of Liver Disease in Pediatric Populations with Cystic Fibrosis
Published on: August 29, 2025
Related Concept Videos
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Myocarditis II: Clinical Features and Diagnostic Tests
Pericarditis II: Clinical Features and Diagnostic Tests
Atherosclerosis II: Clinical Manifestations and Diagnostic Tests
Mitral Stenosis II: Clinical features and Diagnostic Tests
Aortic Regurgitation II: Clinical Features and Diagnostic Tests