Comparison of the Rotational Stability of Two Toric Intraocular Lenses in 1273 Consecutive Eyes
1Altos Eye Physicians, Los Altos, California.
Purpose:
To compare the rotational stability of the 2 most commonly used toric intraocular lenses (TIOLs).
Design:
Retrospective cohort study in a single private practice.
Subjects:
The study included all patients receiving an Acrysof (n = 626) or Tecnis TIOL (n = 647) over an 18-month period from April 2015 to September 2016. Patients were only excluded if their surgery could not be performed using a digital marking system.
Methods:
All patients had cataract surgery performed in the same surgical center with a similar technique. A digital marking system with limbal vessel registration was used to record the axis of the TIOL at the conclusion of surgery. A dilated examination was performed either later on the day of surgery or the next morning, and the postoperative rotation of the 2 TIOL models was compared. Patients who required a return to the operating room for TIOL repositioning were examined to determine risk factors for reoperation and subsequent outcomes.
Main Outcome Measures:
The primary outcome measure was the percentage of eyes with TIOL rotation >5 and >10 degrees. The second main outcome was likelihood of requiring return to the operating room to reposition a rotated TIOL.
Results:
The Acrysof TIOL was less likely to rotate postoperatively, with 91.9% of eyes rotated ≤5 degrees at the first postoperative check compared with 81.8% of Tecnis TIOL eyes (P < 0.0001). This difference persisted for rotation ≤10 degrees (97.8% Acrysof vs. 93.2% Tecnis, P = 0.0002) and ≤15 degrees (98.6% Acrysof vs. 96.4% Tecnis, P = 0.02). The mean rotation was 2.72 degrees (95% confidence interval 2.35-3.08 degrees) for Acrysof and 3.79 degrees (95% confidence interval 3.36-4.22 degrees) for Tecnis TIOLs (P < 0.05). The Tecnis TIOL showed a strong predisposition to rotate counterclockwise, unlike the Acrysof. More Tecnis TIOL patients required repositioning (3.1% vs. 1.6%), but this did not reach statistical significance (P = 0.10). Refractive outcomes were similar between the 2 groups.
Conclusions:
The Acrysof TIOL showed significantly greater rotational stability than the Tecnis TIOL.
More Related Videos
Related Concept Videos
Nuclear Stability
To hold positively charged protons together...
RNA Stability
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Stability
The stability of an LTI system is determined by the roots of its characteristic equation, known as poles. A system is stable if it produces a bounded...
Kinematic Equations for Rotation
For instance, imagine a point A on a rigid body engaged in circular motion. The translational velocity of this particular point can be calculated by taking the time derivatives of the displacement equation, which essentially measures the...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...