A high-flow nasal cannula system with relatively low flow effectively washes out CO2 from the anatomical dead space
Yu Onodera1, Ryo Akimoto2, Hiroto Suzuki2
1Department of Anesthesiology, Faculty of Medicine, Yamagata University, 2-2-2 Iidanishi, Yamagata City, Yamagata Prefecture, 990-9585, Japan. yu1983624@yahoo.co.jp.
Background:
Although clinical studies of the high-flow nasal cannula (HFNC) and its effect on positive end-expiratory pressure (PEEP) have been done, the washout effect has not been well evaluated. Therefore, we made an experimental respiratory model to evaluate the respiratory physiological effect of HFNC.
Methods:
An airway model was made by a 3D printer using the craniocervical 3D-CT data of a healthy 32-year-old male. CO2 was infused into four respiratory lung models (normal-lung, open- and closed-mouth models; restrictive- and obstructive-lung, open-mouth models) to maintain the partial pressure of end-tidal CO2 (PETCO2) at 40 mmHg. HFNC flow was changed from 10 to 60 L/min. Capnograms were recorded at the upper pharynx, oral cavity, subglottic, and inlet sites of each lung model.
Results:
With the normal-lung, open-mouth model, 10 L/min of HFNC flow decreased the subglottic PETCO2 to 30 mmHg. Increasing the HFNC flow did not further decrease the subglottic PETCO2. With the normal-lung, closed-mouth model, HFNC flow of 40 L/min was required to decrease the PETCO2 at all sites. Subglottic PETCO2 reached 30 mmHg with an HFNC flow of 60 L/min. In the obstructive-lung, open-mouth model, PETCO2 at all sites had the same trend as in the normal-lung, open-mouth model. In the restrictive-lung, open-mouth model, 20 L/min of HFNC flow decreased the subglottic PETCO2 to 25 mmHg, and it did not decrease further. As HFNC flow was increased, PEEP up to 7 cmH2O was gradually generated in the open-mouth models and up to 17 cmH2O in the normal-lung, closed-mouth model.
Conclusions:
The washout effect of the HFNC was effective with relatively low flow in the open-mouth models. The closed-mouth model needed more flow to generate a washout effect. Therefore, HFNC flow should be considered based on the need for the washout effect or PEEP.
More Related Videos
09:50Impact Assessment of Repeated Exposure of Organotypic 3D Bronchial and Nasal Tissue Culture Models to Whole Cigarette Smoke
Published on: February 12, 2015
07:41Three-dimensional Cell Culture Model for Measuring the Effects of Interstitial Fluid Flow on Tumor Cell Invasion
Published on: July 25, 2012
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Gene Flow
Flow Cytometry
In...
Flow Sheet
Here's a closer look at the examples of flowsheets commonly used by nurses:
Graphic Sheet Documentation:
Flow Table Test
Concrete is placed within a truncated cone mold that is 8 inches high with an 8-inch base diameter and a 5-inch top diameter. The...