Purpose:
A minimally invasive, partial-assist, intra-atrial blood pump has been proposed, which would unload the left ventricle with a flow path from the left atrium to the arterial system. Flow modulation is a common strategy for ensuring washout in the pump, but it can increase power consumption because it is typically achieved through motor-speed variation. However, if a pump's performance curve had the proper gradient, flow modulation could be realized passively. To achieve this goal, we propose a pump performance operating curve as an alternative to the more standard operating point.
Methods And Results:
Mean-line theory was employed to generate an initial set of geometries that were then tested on a hydraulic test rig at ~20,000 r/min. Experimental results show that the intra-atrial blood pump performed below the operating region; however, it was determined that smaller hub diameter and longer chord length bring the performance of the intra-atrial blood pump device closer to the operating curve.
Conclusion:
We found that it is possible to shape the pump performance curve for specifically targeted gradients over the operating region through geometric variations inside the pump.
More Related Videos
06:10Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
08:34Technique of Minimally Invasive Transverse Aortic Constriction in Mice for Induction of Left Ventricular Hypertrophy
Published on: September 25, 2017
Related Concept Videos
Predicting Molecular Geometry
Coordination Number and Geometry
Design Example: Flow Through a Fire Extinguisher
The key to understanding how the...
Geometry of Hyperbolas
Design Example: Flow of Oil Through Circular Pipes
Internal Energy