Six-month Longitudinal Comparison of a Portable Tablet Perimeter With the Humphrey Field Analyzer
Selwyn Marc Prea1, Yu Xiang George Kong2, Aditi Mehta3
1Department of Optometry & Vision Sciences, The University of Melbourne, Parkville, Victoria, Australia.
Purpose:
To establish the medium-term repeatability of the iPad perimetry app Melbourne Rapid Fields (MRF) compared to Humphrey Field Analyzer (HFA) 24-2 SITA-standard and SITA-fast programs.
Design:
Multicenter longitudinal observational clinical study.
Methods:
Sixty patients (stable glaucoma/ocular hypertension/glaucoma suspects) were recruited into a 6-month longitudinal clinical study with visits planned at baseline and at 2, 4, and 6 months. At each visit patients undertook visual field assessment using the MRF perimetry application and either HFA SITA-fast (n = 21) or SITA-standard (n = 39). The primary outcome measure was the association and repeatability of mean deviation (MD) for the MRF and HFA tests. Secondary measures were the point-wise threshold and repeatability for each test, as well as test time.
Results:
MRF was similar to SITA-fast in speed and significantly faster than SITA-standard (MRF 4.6 ± 0.1 minutes vs SITA-fast 4.3 ± 0.2 minutes vs SITA-standard 6.2 ± 0.1 minutes, P < .001). Intraclass correlation coefficients (ICC) between MRF and SITA-fast for MD at the 4 visits ranged from 0.71 to 0.88. ICC values between MRF and SITA-standard for MD ranged from 0.81 to 0.90. Repeatability of MRF MD outcomes was excellent, with ICC for baseline and the 6-month visit being 0.98 (95% confidence interval: 0.96-0.99). In comparison, ICC at 6-month retest for SITA-fast was 0.95 and SITA-standard 0.93. Fewer points changed with the MRF, although for those that did, the MRF gave greater point-wise variability than did the SITA tests.
Conclusions:
MRF correlated strongly with HFA across 4 visits over a 6-month period, and has good test-retest reliability. MRF is suitable for monitoring visual fields in settings where conventional perimetry is not readily accessible.
Related Concept Videos
Longitudinal Research
Longitudinal Studies
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...