The intracavitary electrocardiography method for positioning the tip of epicutaneous cava catheter in neonates: Pilot
Antonella Capasso1, Rossella Mastroianni1, Annalisa Passariello2,3
11 Neonatal Intensive Care Unit, Hospital Sant'Anna and San Sebastiano, Caserta, Italy.
Purpose::
The neonatologists of Sant'Anna and San Sebastiano Hospital of Caserta have carried out a pilot study investigating the safety, feasibility, and accuracy of intracavitary electrocardiography for neonatal epicutaneous cava catheter tip positioning.
Patients And Methods::
We enrolled 39 neonates (1-28 days of postnatal age or correct age lower than 41 weeks) requiring epicutaneous cava catheter in the district of superior vena cava (head-neck or upper limbs). Intracavitary electrocardiography was applicable in 38 neonates.
Results::
No significant complications related to intracavitary electrocardiography occurred in the studied neonates. The increase in P wave on intracavitary electrocardiography was detected in 30 cases. Of the remaining eight cases, six malpositioned catheters tipped out of cavoatrial junction-target zone (chest x-ray and echocardiographical control) and two were false negative (tip located in target zone). The match between intracavitary electrocardiography and x-ray was observed in 29/38 cases, and the same ratio between intracavitary electrocardiography and echocardiography was detected.
Conclusion::
We conclude that the intracavitary electrocardiography method is safe and accurate in neonates as demonstrated in pediatric and adult patients. The applicability of the method is 97% and its feasibility is 79%. The overall accuracy is 76% but it rises to 97% if "peak" P wave is detected.