Related Experiment Video
Updated: Feb 13, 2026

In Vivo Evaluation of the Mechanical and Viscoelastic Properties of the Rat Tongue
Published on: July 6, 2017
Assessment of the viscoelastic mechanical properties of polycarbonate urethane for medical devices
Agnes Beckmann1, Yousef Heider1, Marcus Stoffel1
1Institute of General Mechanics, RWTH Aachen University, Templergraben 64, D-52062 Aachen, Germany.
Abstract:
The underlying research work introduces a study of the mechanical properties of polycarbonate urethane (PCU), used in the construction of various medical devices. This comprises the discussion of a suitable material model, the application of elemental experiments to identify the related parameters and the numerical simulation of the applied experiments in order to calibrate and validate the mathematical model. In particular, the model of choice for the simulation of PCU response is the non-linear viscoelastic Bergström-Boyce material model, applied in the finite-element (FE) package Abaqus®. For the parameter identification, uniaxial tension and unconfined compression tests under in-laboratory physiological conditions were carried out. The geometry of the samples together with the applied loadings were simulated in Abaqus®, to insure the suitability of the modelling approach. The obtained parameters show a very good agreement between the numerical and the experimental results.
More Related Videos
Related Concept Videos
Nephrotic Syndrome II : Assessment and Medical Management
Inhaled Medications
Physical and Chemical Properties of Matter
Properties of Transition Metals
The Quantum-Mechanical Model of an Atom
Endocarditis III: Medical Management