Related Experiment Video
Updated: Feb 13, 2026

Fabrication of Small Caliber Stent-grafts Using Electrospinning and Balloon Expandable Bare Metal Stents
Published on: October 26, 2016
Longitudinal Deformation of Distal Edge in a New-Generation Stent Caused by Guidewire Entrapment
Abstract:
Longitudinal stent deformation, described in some older stent geometries, prompted design modifications such as reinforcing struts on the proximal end. However, distal edges of stents-also subject to longitudinal force-have not been reinforced. We report a case of guidewire entrapment that deformed the distal edge of a new-generation stent during percutaneous coronary intervention, and we describe our efforts to restore the stent to its initial length. This case highlights the risk of manipulating equipment beyond the position of a newly deployed stent, the ongoing potential for deformation of distal edges in newer stent platforms, and the advisability of treating distal lesions before proximal ones.
More Related Videos
Related Concept Videos
Longitudinal Research
Plastic Deformations
Plastic Deformations
Longitudinal Studies
Temperature Dependent Deformation
Deformations in a Symmetric Member in Bending
When the member is segmented into tiny cubic elements, it is observed that the primary stress...