Accuracy of an oscillometric blood pressure monitor in anesthetized pigs
Related Experiment Video
Updated: Feb 13, 2026

Non-Terminal Blood Sampling Techniques in Guinea Pigs
Published on: October 11, 2014
Rachel Reed1, Michele Barletta1, Janet Grimes2
11 Department of Large Animal Medicine, College of Veterinary Medicine, University of Georgia, USA.
Abstract:
The purpose of this study is to evaluate the accuracy of an oscillometric blood pressure monitor in anesthetized pigs. Invasive blood pressure (IBP) and noninvasive blood pressure (NIBP) measurements were taken using a DRE Waveline Pro multiparameter monitor at four different time points in 17 pigs undergoing injectable anesthesia. NIBP measurements were taken on both the thoracic and pelvic limbs. Bland Altman analysis was used to assess agreement between methods and a linear mixed-effects model was used to evaluate the effect of cuff position and blood pressure on bias. Invasive systolic arterial pressure (SAP) ranged between 112 and 161 mmHg (mean ± SD: 138.8 ± 13.3; median: 139.5). Invasive diastolic arterial pressure (DAP) ranged between 60 and 104 mmHg (mean ± SD: 86.0 ± 9.1; median: 87.0). Invasive mean arterial pressure (MAP) ranged between 79 and 121 mmHg (mean ± SD: 103.2 ± 9.3; median 103.0). Only the diastolic and mean measurements obtained from the pelvic limb met criteria outlined by the American College of Internal Medicine for required accuracy of NIBP monitors. Bias was significantly higher in the thoracic limb in comparison to the pelvic limb and was significantly higher at blood pressures above median. In general, NIBP measurements underestimated IBP measurements. In conclusion, the use of the DRE Waveline Pro to assess NIBP in anesthetized pigs may be useful in monitoring trends in mean and diastolic blood pressure and is most accurate when used on the pelvic limb.
More Related Videos
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Blood Pressure
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the...
Measurement of Blood Pressure
Factors affecting Blood pressure
Physiological Factors:
Alterations in Blood Pressure
Hypertension (High blood pressure)
Hypertension occurs when blood pressure readings consistently exceed the normal range. It is diagnosed when systolic blood pressure (the top number, indicating pressure while the heart...