Olbert dilatation catheter system: complication due to inexpert handling
Cardiovascular and Interventional Radiology
|January 1, 1987
Summary
Inexpert handling of the Olbert catheter system can cause kinking. An improvised technique was used to successfully remove the catheter, highlighting the need for proper training.
Area of Science:
- Medical Devices
- Interventional Radiology
- Urology
Background:
- The Olbert catheter system is utilized in various interventional procedures.
- Proper handling is crucial to prevent device-related complications.
Observation:
- A complication occurred during Olbert catheter system use due to a closed proximal clamp during balloon inflation.
- This led to distal catheter kinking within the inflated balloon.
Findings:
- The kinking caused significant difficulty in catheter removal.
- An improvised coaxial sheath technique was successfully employed to overcome the removal obstruction.
Implications:
- This case underscores the importance of meticulous technique and training for the Olbert catheter system.
- The improvised coaxial sheath technique offers a potential solution for similar catheter impaction scenarios.