Flow variation as a factor determining repeatability of the internal standard-based qualitative and quantitative
Paweł Mateusz Nowak1, Michał Woźniakiewicz1, Paweł Kościelniak1
1Jagiellonian University in Kraków, Faculty of Chemistry, Department of Analytical Chemistry, Gronostajowa St. 2, 30-387 Kraków, Poland.
Abstract:
The use of migration times and peak areas referred to another sample component - internal standard, brings many benefits in improving reliability of capillary electrophoresis. However, it is quite commonly overlooked that despite relative migration time and peak area ratio are more stable than the absolute values upon alteration in the flow rate, some shift should always be expected. The present work offers a new look at this analytically-important issue. We have derived a simple model allowing to estimate the magnitude of error for the selected pair of molecules of known mobilities upon the given flow alteration. Then, we have confronted the theoretical predictions with the experimental results obtained for the model sample separated in various flow conditions reached by the external pressure manipulation, including several internal standards of different mobilities. A good agreement has been obtained, pointing out that the magnitude of error may be large even for the seemingly "good" internal standards. Several potentially useful means have been tested to address this issue: the use of electrophoretic mobilities and electrophoretic mobility ratios instead migration times in the qualitative analysis, and performing time-correction of peak area ratios, or alternatively, transformation of electropherograms from the time-related scale into the electrophoretic mobility-related scale in the quantitative analysis. We have also considered some additional factors. The results may be of interest for all users dealing with the development and optimization of analytical methods using capillary electrophoresis.
More Related Videos
14:12Amplification of Escherichia coli in a Continuous-Flow-PCR Microfluidic Chip and Its Detection with a Capillary Electrophoresis System
Published on: November 21, 2023
10:33Multi-locus Variable-number Tandem-repeat Analysis of the Fish-pathogenic Bacterium Yersinia ruckeri by Multiplex PCR and Capillary Electrophoresis
Published on: June 17, 2019
Related Concept Videos
Capillary Electrophoresis: Instrumentation
Variation: Normal Distribution, Range, and Standard Deviation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Qualitative Analysis
For instance, group IV...
Qualitative Analysis
There are two main approaches to qualitative analysis:...
What is Variation?
The range, standard deviation, standard error, and variance are the different measures of variation.
Range: The range is the difference between its maximum and...