Cardiac Output Monitoring by Pulse Contour Analysis, the Technical Basics of Less-Invasive Techniques
1Department of Intensive Care Medicine, Center of Anesthesiology and Intensive Care Medicine, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
Abstract:
Routine use of cardiac output (CO) monitoring became available with the introduction of the pulmonary artery catheter into clinical practice. Since then, several systems have been developed that allow for a less-invasive CO monitoring. The so-called "non-calibrated pulse contour systems" (PCS) estimate CO based on pulse contour analysis of the arterial waveform, as determined by means of an arterial catheter without additional calibration. The transformation of the arterial waveform signal as a pressure measurement to a CO as a volume per time parameter requires a concise knowledge of the dynamic characteristics of the arterial vasculature. These characteristics cannot be measured non-invasively and must be estimated. Of the four commercially available systems, three use internal databases or nomograms based on patients' demographic parameters and one uses a complex calculation to derive the necessary parameters from small oscillations of the arterial waveform that change with altered arterial dynamic characteristics. The operator must ensure that the arterial waveform is neither over- nor under-dampened. A fast-flush test of the catheter-transducer system allows for the evaluation of the dynamic response characteristics of the system and its dampening characteristics. Limitations to PCS must be acknowledged, i.e., in intra-aortic balloon-pump therapy or in states of low- or high-systemic vascular resistance where the accuracy is limited. Nevertheless, it has been shown that a perioperative algorithm-based use of PCS may reduce complications. When considering the method of operation and the limitations, the PCS are a helpful component in the armamentarium of the critical care physician.
More Related Videos
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Voltammetric Techniques: Pulse Voltammetry