A new approach for reducing beam hardening artifacts in polychromatic X-ray computed tomography using more accurate
Hui Wang1, Yanan Xu1, Hongli Shi1
1School of Biomedical Engineering, Capital Medical University, Beijing, China.
Purpose:
Metal artifacts severely degrade CT image quality in clinical diagnosis, which are difficult to removed, especially for the beam hardening artifacts. The metal artifact reduction (MAR) based on prior images are the most frequently-used methods. However, there exists a lot misclassification in most prior images caused by absence of prior information such as the spectrum distribution of X-ray beam source, especially many or big metal included. The purpose of this work is to find a more accurate prior image to improve image quality.
Methods:
The proposed method comprise of following four steps. First, the metal image is segmented by thresholding an initial image, where the metal traces are identified in the initial projection data using the forward projection of the metal image. Second, the accurate absorbent model of certain metal image is calculated according to the spectrum distribution of certain X-ray beam source and energy-dependent attenuation coefficients of metal. Then, a new metal image is reconstructed by the general analytical reconstruction algorithm such as filtered back projection (FPB). The prior image is obtained by segmenting the difference image between the initial image and the new metal image into air, tissue and bone. Finally, the initial projection data are normalized by dividing the projection data of prior image pixel to pixel, the corrected image is obtained by interpolation, denormalization and reconstruction.
Results:
Some clinical images with dental fillings and knee prostheses are used to evaluate the proposed algorithm and normalized metal artifact reduction (NMAR) and linear interpolation (LI) method. The results demonstrate the artifacts can be reduced efficiently by the proposed method.
Conclusions:
The proposed method could obtain an exact prior image using the prior information about X-ray beam source and energy-dependent attenuation coefficients of metal. As a result, the better performance of reducing beam hardening artifacts can be improved, even though there were many or big implants. Moreover, the process of the proposed method is rather simple and little extra calculation burden is necessary. It has superiorities over other algorithms when include big or many implants.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...