Postnatal dilatation of umbilical cord vessels and its impact on wall integrity: Prerequisite for the artificial
Jenny Peng1,2, Niels Rochow1, Mohammadhossein Dabaghi3
11 Division of Neonatology, Department of Pediatrics, McMaster University, Hamilton, ON, Canada.
Introduction:
A lung assist device, which acts as an artificial placenta, can provide additional gas exchange for preterm and term newborns with respiratory failure. The concept of the lung assist device requires a large bore access via umbilical vessels to allow pumpless extracorporeal blood flow rates up to 30 mL/kg/min. After birth, constricted umbilical vessels need to be reopened for vascular access. The objective is to study the impact of umbilical vessel expansion on vessel integrity for achieving large bore access.
Methods:
Umbilical cords from healthy term deliveries were cannulated and dilatated with percutaneous transluminal angioplasty catheters in 1 mm increments from 4 to 8 mm for umbilical artery and from 4 to 15 mm for umbilical vein, n = 6 per expansion diameter. Paraffin-embedded transverse sections of dilated and control samples were HE & Van Gieson stained. Effects of dilatation, shown by splitting, were measured.
Results:
Umbilical vessel expansion led to concentric splitting, shown by areas devoid of extracellular matrix and nuclei in the tunica intima and media. No radial splitting was observed. Results suggest an expansion threshold of umbilical artery at 6 mm and umbilical vein at 7 mm, while maximal splitting was observed above this threshold (3.6 ± 0.8%, p = 0.043 for umbilical artery 7 mm and 6.3 ± 1.8%, p = 0.048 for umbilical vein 8 mm). Endothelial cell sloughing was present in all dilated samples but not in the control samples.
Conclusion:
The suggested thresholds for safe expansions are similar to in utero umbilical vessel diameters and demonstrate a proof of concept for attaining large bore access for the lung assist device.
More Related Videos
12:17The 4-vessel Sampling Approach to Integrative Studies of Human Placental Physiology In Vivo
Published on: August 2, 2017
09:29Clinical Application of Sleeping Beauty and Artificial Antigen Presenting Cells to Genetically Modify T Cells from Peripheral and Umbilical Cord Blood
Published on: February 1, 2013
Related Concept Videos
Plant Cell Wall
Cardiomyopathy II: Dilated Cardiomyopathy
Impact of Groups on Groups
Impact
When particles with different initial velocities collide, they induce deformation by applying equal and opposite impulses. At the point of maximum deformation, the particles move together with...
Integration by Parts: Indefinite Integrals
Integration by Parts: Definite Integrals