FEASIBILITY STUDY OF USING PCXMC 2.0 TO ESTIMATE PATIENT DOSE ARISING FROM DEXA SCANS
Hollie Harmer1, Sally Radford1, Mary Smail1
1Department of Medical Physics and Bioengineering, Bristol Royal Infirmary, University Hospitals Bristol NHS Foundation Trust, Bristol, UK.
Abstract:
Patients undergoing dual energy X-ray absorption (DEXA) scans are exposed to small doses of ionizing radiation. Few papers have been published on the effective dose and organ dose for patients undergoing such scans on newer DEXA scanners. PCXMC 2.0 was used to calculate adult patient dose arising from DEXA scans. PCXMC 2.0 calculations were compared to published data and a discrepancy was noted for organ dose. Following this, effective and organ dose were measured on an anthropomorphic phantom using TLDs as a second comparison. The mean dose from 50 scans (minus background radiation) was measured. The dose calculated from PCXMC 2.0 compared to published data shows very good agreement for effective dose but a difference for organ dose. Our TLD data and PCXMC 2.0 data for organ dose have a closer agreement, within 20%. We are confident in using PCXMC 2.0 to calculate adult patient effective and organ dose arising from DEXA scans.
Related Concept Videos
Drug Dosing: Geriatric Patients
Drug Dosing: Obese Patients
Bioavailability Study Design: Single Versus Multiple Dose Studies
Leaky Scanning
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
What are Estimates?
The estimate for the mean of a sample is denoted by ͞x, whereas the mean of the population is designated as μ. Further, parameters such...