Comparative evaluation of Scheimpflug tomography parameters between thin non-keratoconic, subclinical keratoconic,
Samira Huseynli1, José Salgado-Borges2, Jorge L Alio3
11 National Ophthalmology Center named after academician Zarifa Aliyeva, Baku, Azerbaijan.
Purpose:
To evaluate and compare the topographic and topometric parameters, thickness profile data, and data from enhanced elevation maps of thin non-keratoconic, subclinical keratoconic, and mild keratoconic corneas with the Pentacam Scheimpflug corneal tomography and to study the usefulness of different parameters to differentiate keratoconus from topographically normal thin corneas.
Methods:
The study included 30 eyes with subclinical keratoconus, 30 eyes with mild-stage keratoconus, and 54 healthy eyes with minimal pachymetry ≤500 µm, with a mean age of 21.19 ± 2.97, 21.75 ± 1.93, and 21.5 ± 2.95 years, respectively. The area under the receiver operating characteristic curves was used to analyze the diagnostic significance of the Pentacam parameters.
Results:
The anterior and posterior corneal elevations, pachymetric progression, the percentage of thickness increase measurements, overall D value, and topometric indices were statistically significantly higher in subclinical and mild keratoconic corneas than in normal eyes with thin cornea (p < 0.05). All these parameters had sufficient strength (area under the receiver operating characteristic curves >0.90) to differentiate clinical keratoconus. Posterior elevation showed the excellent area under the receiver operating characteristic curves with 100% sensitivity and 100% specificity for this purpose. However, among all parameters studied, the anterior elevation (0.935) showed the excellent area under the receiver operating characteristic curves to differentiate subclinical keratoconus, followed by posterior elevation (0.897), index of height decentration (0.887), and D value (0.882).
Conclusion:
The parameters derived from the Scheimpflug device, such as corneal elevations and overall D value, can effectively differentiate subclinical and clinical keratoconus from non-keratoconic thin cornea eyes. However, the specificity levels of these parameters were relatively limited in the diagnosis of subclinical keratoconus.
Related Concept Videos
Wave Parameters
Comparing Copy Number Variations and SNPs
Copy number variations or CNVs are the structural variations that cover more than 1kb of DNA sequence. The single nucleotide polymorphism (SNP), on the other hand, is a single nucleotide change or a point mutation that is found in more than 1%...
Comparative Excretory Systems
Comparing Experimental Results: Student's t-Test
Comparing the Survival Analysis of Two or More Groups
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...