Related Experiment Video
Updated: Feb 12, 2026

Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Transcatheter Pulmonary Valve Replacement With the Melody Valve in Small Diameter Expandable Right Ventricular
Shabana Shahanavaz1, Athar M Qureshi2, Daniel S Levi3
1Division of Cardiology, Department of Pediatrics, Washington University in St. Louis School of Medicine, St. Louis, Missouri.
Objectives:
This study sought to evaluate the safety, feasibility, and outcomes of transcatheter pulmonary valve replacement (TPVR) in conduits ≤16 mm in diameter.
Background:
The Melody valve (Medtronic, Minneapolis, Minnesota) is approved for the treatment of dysfunctional right ventricular outflow tract (RVOT) conduits ≥16 mm in diameter at the time of implant. Limited data are available regarding the use of this device in smaller conduits.
Methods:
The study retrospectively evaluated patients from 9 centers who underwent percutaneous TPVR into a conduit that was ≤16 mm in diameter at the time of implant, and reported procedural characteristics and outcomes.
Results:
A total of 140 patients were included and 117 patients (78%; median age and weight 11 years of age and 35 kg, respectively) underwent successful TPVR. The median original conduit diameter was 15 (range: 9 to 16) mm, and the median narrowest conduit diameter was 11 (range: 4 to 23) mm. Conduits were enlarged to a median diameter of 19 mm (29% larger than the implanted diameter), with no difference between conduits. There was significant hemodynamic improvement post-implant, with a residual peak RVOT pressure gradient of 7 mm Hg (p < 0.001) and no significant pulmonary regurgitation. During a median follow-up of 2.0 years, freedom from RVOT reintervention was 97% and 89% at 2 and 4 years, respectively, and there were no deaths and 5 cases of endocarditis (incidence rate 2.0% per patient-year).
Conclusions:
In this preliminary experience, TPVR with the Melody valve into expandable small diameter conduits was feasible and safe, with favorable early and long-term procedural and hemodynamic outcomes.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...