A comprehensive evaluation of angular range and separation on image quality, image registration, and imaging dose for
1National Cancer Center/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, 100021, China.
Abstract:
We aimed to comprehensively evaluate the quality, dose, and registration of cone beam computed tomography (CBCT) images acquired with different angular range (θ) and angular separation (Δθ). Elekta XVI volumetric imaging system was used in this study. CBCT images of a Catphan 503 phantom were reconstructed with projections that were acquired under 17 × 8 different combinations of θ (200° to 360°) and Δθ (0.27° to 3.24°). The evaluation indices were contrast-to-noise ratio (CNR), spatial resolution (modulation transfer function [MTF50%]), uniformity, and registration accuracy. The weighted computed tomography dose index (CTDIw) of CTDI Head dosimetry phantom was measured for estimating radiation exposure. CNR changed little from θ = 200° to 260°, but increased by ≈ 2.0 times from θ = 260° to 360°. CNR also increased by ≈ 2.5 times from Δθ = 2.16° to 0.27°. MTF50% increased with Δθ being decreased, but had no dependence on θ. Image uniformity became better when θ was smaller or larger than 260°, but there was no correlation with Δθ. Registration of the overall body had a maximum error of 0.3 mm. The error of low-contrast objects' registration could reach as much as 1.3 mm. CTDIw showed a linear dependence on θ and 1/Δθ. Although the imaging dose is reduced with θ being decreased or Δθ being increased, the image quality may be degraded. Accuracy of image registration of a rigid body can be achieved; however, low-contrast objects' registration using images with inferior quality will introduce unacceptable errors for clinical use. Further, the benefits and extra dose received by the patients should be balanced, and scan protocols with θ between 210° and 260° or Δθ smaller than 0.54° should not be recommended.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
13:10Combined Near-infrared Fluorescent Imaging and Micro-computed Tomography for Directly Visualizing Cerebral Thromboemboli
Published on: September 25, 2016
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies VII: Vascular Imaging
X-ray Imaging
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...