Evaluation and Comparison of Hemodynamic Performance of Three ECLS Systems in a Simulated Adult Cardiogenic Shock
Akif Ündar1,2, Shigang Wang1, Morgan Moroi1
1Department of Pediatrics, Penn State Health Pediatric Cardiovascular Research Center, Penn State College of Medicine, Penn State Health Children's Hospital, Hershey, PA, USA.
Abstract:
The objective of this study was to evaluate three commercially available ECLS systems with rotary pumps in terms of circuit pressure, pressure drop, perfusion modes, and hemodynamic energy transmission in a simulated adult cardiogenic shock model. One circuit consisted of a Cardiohelp system, which included a Cardiohelp console and HLS Module Advanced 7.0 tubing set with integrated centrifugal pump and oxygenator. The alternative circuit was composed of a Quadrox-D Adult oxygenator connected in series with either an i-cor diagonal pump and console or a Rotaflow centrifugal pump and console. The circuit was primed with lactated Ringer's solution and packed red blood cells (hematocrit 40%). The trials were conducted at flow rates of 1-5 L/min with pseudo patient pressures of 60 mm Hg and 80 mm Hg. Pulsatile flow was tested when using the i-cor system. Mean pre-oxygenator pressure and pressure drop across ECLS circuit (including oxygenator and arterial tubing) were lower when using the Cardiohelp system as compared to the Rotaflow and i-cor systems (P < 0.01). The i-cor system was able to deliver more hemodynamic energy to the pseudo patient because of its ability to produce pulsatile flow (P < 0.01). The Cardiohelp HLS Module Advanced 7.0 integrated oxygenator had a lower resistance than the Quadrox-D oxygenator. Although the compact Cardiohelp system had a better hemodynamic performance when compared to Rotaflow and i-cor systems, the pulsatile flow of the i-cor system delivered more hemodynamic energy to the pseudo patient. This may render more physiological benefits in high-risk patients on ECLS.
More Related Videos
Related Concept Videos
Pharmacokinetic Models: Comparison and Selection Criterion
Physiological models take a detailed approach by considering specific molecular processes. They can predict drug distribution, metabolism, and elimination changes, providing a comprehensive understanding of how drugs interact with the body.
Shock Waves
When the source's speed approaches the speed of sound, constructive interference between successive wavefronts emitted by the source occurs immediately behind it. Initially, scientists believed that this constructive interference would result in such high...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Pharmacodynamic Models: Link Model and Systems Pharmacodynamic Model
Self-Evaluation Maintenance Model
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...