Influence of room temperature on magnesium isotope measurements by multi-collector inductively coupled plasma mass
Xing-Chao Zhang1, An-Yu Zhang2, Zhao-Feng Zhang3
1CAS Key Laboratory of Crust-Mantle Materials and Environments, Department of Earth and Space Sciences, University of Science and Technology of China, Hefei, 230026, China.
Rationale:
We observed that the accuracy and precision of magnesium (Mg) isotope analyses could be affected if the room temperature oscillated during measurements. To achieve high-quality Mg isotopic data, it is critical to evaluate how the unstable room temperature affects Mg isotope measurements by multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS).
Methods:
We measured the Mg isotopes for the reference material DSM-3 using MC-ICP-MS under oscillating room temperatures in spring. For a comparison, we also measured the Mg isotopes under stable room temperatures, which were achieved by the installation of an improved temperature control system in the laboratory.
Results:
The δ26 Mg values measured under oscillating room temperatures have a larger deviation (δ26 Mg from -0.09 to 0.08‰, with average δ26 Mg = 0.00 ± 0.08‰) than those measured under a stable room temperature (δ26 Mg from -0.03 to 0.03‰, with average δ26 Mg = 0.00 ± 0.02‰) using the same MC-ICP-MS system.
Conclusions:
The room temperature variation can influence the stability of MC-ICP-MS. Therefore, it is critical to keep the room temperature stable to acquire high-precision and accurate isotopic data when using MC-ICP-MS, especially when using the sample-standard bracketing (SSB) correction method.
More Related Videos
09:05Imaging Metals in Brain Tissue by Laser Ablation - Inductively Coupled Plasma - Mass Spectrometry LA-ICP-MS
Published on: January 22, 2017
07:25Quantitative Metabolomics of Saccharomyces Cerevisiae Using Liquid Chromatography Coupled with Tandem Mass Spectrometry
Published on: January 5, 2021
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Mass Spectrometry: Isotope Effect
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Isotopes
An element's atomic mass, or weight,...