Related Experiment Video
Updated: Feb 12, 2026

Microsurgical Creation of Giant Bifurcation Aneurysms in Rabbits for the Evaluation of Endovascular Devices
Published on: September 8, 2023
Endobag separation - an ominous sign after endovascular aneurysm sealing with the Nellix device
Sadia Tasleem1, Hassan Badri1, Narlawar Ranjeet2
11 Department of Vascular and Endovascular Surgery, The Royal Oldham Hospital, Pennine Acute Hospitals NHS Trust, Manchester, United Kingdom.
Abstract:
Our report aims to alert physicians performing endovascular aneurysm sealing with the Nellix endoprosthesis to the risk of rupture when endobag separation is noticed on computed tomography. Endobag separation when observed with acute presentation of abdominal pain is an indicator of imminent rupture and prompt interventional treatment should be undertaken.