Increasing robustness of radial GRASE acquisition for SAR-reduced brain imaging
Melisa Okanovic1, Michael Völker1, Robert Trampel2
1Department of Experimental Physics 5, University of Würzburg, Würzburg, Germany.
Purpose:
To improve a radial multi-slice 2D gradient- and spin-echo (GRASE) sequence and provide an appropriate image reconstruction technique for SAR-reduced high-resolution neuroimaging.
Methods:
Additional readout gradients per radio-frequency (RF) refocusing allow for a reduced number of RF pulses. In this way, a specific absorption rate (SAR) reduction is achieved and the application at high-field systems becomes more feasible. A phase insensitive image reconstruction is proposed to reduce signal dropout artifacts originating from opposite readout polarities. In addition, the image reconstruction allows for the calculation of images with varying contrast from one measurement.
Results:
Results obtained at 3T and 7T demonstrate a SAR-reduction of at least 66% for a single-slice experiment with radial GRASE. The reduced SAR is used for an increased spatial coverage without increasing the measurement time. Experiments at 3T and 7T showed that the visual image quality is comparable to standard TSE and GRASE sequences with the same measurement parameters. Using higher EPI factors and the presented image reconstruction, artifact-free images with a significant SAR-reduction can be achieved.
Conclusion:
Radial GRASE enables SAR-reduced acquisitions of high-resolution brain images with different contrasts from one measurement and is a promising sequence for high-field neuroimaging.
Related Concept Videos
Radial System Protection
In a radial system with a fault downstream of the third breaker, ideally, only the third breaker will open, isolating the fault and interrupting the load connected beyond it. The second breaker has a longer delay setting,...
Light Acquisition
Increasing Function
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Reducing Line Loss
With a step-up transformer at the source, the voltage is increased, thereby reducing the current in the transmission lines since power loss in...