The Performance of a Zirconium-based Root Filling Material with Artifact Reduction Properties in the Detection of
Aaron Fox1, Bettina Basrani1, Ernest W N Lam2
1Graduate Program in Endodontics, Faculty of Dentistry, University of Toronto, Toronto, Ontario, Canada.
Introduction:
Limited field of view cone-beam computed tomographic (CBCT) imaging has been used to augment clinical testing of vertical root fractures (VRFs); however, the presence of gutta-percha (GP) in the canal space generates substantial imaging artifacts that make fracture detection difficult. The purpose of this study was to evaluate the influence of a zirconium (Zr)-based root filling material with radiologic properties that reduce beam hardening (BH) artifacts using CBCT imaging in the in vitro diagnosis of VRFs.
Methods:
One hundred seventy-six single-rooted mandibular premolar teeth were obtained, and half of these teeth were filled with GP or Zr (CPoint; EndoTechnologies, LLC, Shrewsbury, MA). VRFs were induced in 44 decoronated teeth in each group using an Instron (Norwood, MA) Universal Testing Machine. Each root was then placed in a dry human mandible and imaged with the Carestream 9000 3D CBCT system (Carestream Dental, Atlanta, GA). The images were evaluated by 6 oral maxillofacial radiologists (OMRs) and residents.
Results:
The sensitivity was greater for detecting VRFs in the Zr group than the GP group (P = .035). However, the specificity was greater for the GP group than the Zr group (P = .028). Receiver operating characteristic area under the curve values were greater for the Zr group than the GP group, but these differences were not statistically significant. The OMRs outperformed the residents in the detection of VRFs in the Zr group with respect to specificity (P = .006) and positive predictive value (P = .012).
Conclusions:
The reduced BH of the Zr group improved the sensitivity of the detection of artificially induced VRFs. The ability to detect VRFs in the Zr group was further enhanced by clinical experience.
Related Concept Videos
Properties of the Root Locus
To determine if a point lies on the root locus, the criterion involves the sum of angles contributed by all poles and zeros to that point. Specifically, this sum must be an odd multiple of 180 degrees. The gain at any point on...
Primary and Secondary Growth in Roots and Shoots
Root Mean Square
For example, consider the velocity of gas molecules in a container. The gas molecules are moving in different directions, which might impart positive and negative...
Root-Locus Method
This system can be represented by a block...
Construction of Root Locus
For positive gain values, the root locus exists on the real axis to the left of an odd number of finite open-loop poles or zeros. The root locus starts at the open-loop poles and traces the paths of the closed-loop poles as the gain...
Plotting and Calibrating the Root Locus
The maximum gain occurs at the breakaway points between open-loop poles on the real axis, while the minimum gain is...