High altitude arterialised capillary earlobe blood gas measurement using the Abbott i-STAT
Christopher T Lewis1, W L Malein2, I Chesner3
1Academic Foundation Programme, NHS Highland, Inverness, UK.
Introduction:
Measurement of physiological parameters in extreme environments is essential to advancing knowledge, prophylaxis and treatment of altitude sickness. Point-of-care testing facilitates investigation in non-specialist and remote settings, as well as becoming increasingly popular at the bedside for real-time results in the clinical environment. Arterialised capillary earlobe blood gases are recommended as a valid alternative to arterial sampling in research. This study aimed to test the feasibility of obtaining and analysing daily earlobe samples at high altitude.
Methods:
From 17 to 24 January 2016, 24 participants on a research expedition to Ecuador underwent daily earlobe blood gas measurements including pH, partial pressure of oxygen and partial pressure of carbon dioxide to 5043 m. Samples were analysed using an Abbott i-STAT blood gas analyser and G3+ cartridges.
Results:
Daily measurements were successfully obtained and analysed at the point of care in 23/24 participants and were well tolerated with no adverse events. 12% (27/220) cartridges failed and required repeat sampling.
Conclusions:
Daily earlobe blood gas analysis using the Abbott i-STAT is feasible in a protected environment at high altitude. Participants and equipment should be kept warm before and during testing. Spare cartridges should be available. This methodology may be useful for both research and therapeutic measurements in remote, rural and wilderness medicine.
More Related Videos
08:44Author Spotlight: Integrating Alveolar-Capillary Reserve Measurements in Exercise Adaptation and Therapeutic Strategies
Published on: February 2, 2024
07:12Time-resolved Förster Resonance Energy Transfer Assays for Measurement of Endogenous Phosphorylated STAT Proteins in Human Cells
Published on: September 9, 2021
Related Concept Videos
Gas Exchange and Transport
Measurement of Blood Pressure
The JAK-STAT Signaling Pathway
Kinetic Molecular Theory and Gas Laws Explain Properties of Gas Molecules
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer