Related Experiment Video
Updated: Feb 12, 2026

Reverse Total Shoulder Arthroplasty
Published on: July 5, 2011
The Impact of Pinless Navigation in Conventionally Aligned Total Knee Arthroplasty
P Koenen1, M M Schneider1, T R Pfeiffer1
1Department of Orthopaedics, Trauma Surgery and Sports Medicine, Cologne-Merheim Medical Center, University of Witten/Herdecke, Cologne, Germany.
Background:
Restoration of the mechanical axis is a main objective in total knee replacement (TKR). Aim of this study was to analyse the verification tool of a pinless navigation system in conventional TKR (cTKR).
Methods:
In a prospective study, 147 TKR were performed by conventional technique. Using the "pinless verification" mode of a smartphone based navigation system, the cutting block position and final resection plane for distal femur and proximal tibial resection were measured. If necessary, the block position or resection level were optimized, corrections were protocolled. Postoperatively, standardized radiographs were performed.
Results:
In 65.3%, intraoperative measurements changed the surgical procedure (corrections: 20.4% femoral, 25.9% tibial, 19% both). The additional time for surgery compared to cTKR averaged 6 minutes (79 ± 15 versus 73 ± 17 minutes). Using navigation data, the final femoral and tibial axes were in 93% within a range of ±2°. A mean difference of 1.4° and 1.6° could be shown between the final measurement of the navigation system and the postoperative mLDFA and mMPTA.
Conclusion:
Intraoperative pinless navigation has impact on the surgical procedure in the majority of cTKR. It represents a less time-consuming tool to improve implant position while maintaining the routine of conventional technique.
Related Concept Videos
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Sign Convention
The normal force acts perpendicular to the beam's cross-section and can...
Impact of Groups on Groups
Enolate Mechanism Conventions
Impact
When particles with different initial velocities collide, they induce deformation by applying equal and opposite impulses. At the point of maximum deformation, the particles move together with...
Design Example: Alignment of a Road Line Using GIS