Effects of heterogeneities in dose distributions under nonreference conditions: Monte Carlo simulation vs dose
Cristiano Queiroz Melo Reis1, Patricia Nicolucci2, Saulo S Fortes1
1Departamento de Física Médica, Instituto Nacional de Câncer José Alencar Gomes da Silva (INCA), Praça da Cruz Vermelha, Rio de Janeiro, RJ 20230-130, Brazil.
Abstract:
The purpose of this study is to evaluate the performance of dose calculation algorithms used in radiotherapy treatment planning systems (TPSs) in comparison with Monte Carlo (MC) simulations in nonelectronic equilibrium conditions. MC simulations with PENELOPE package were performed for comparison of doses calculated by pencil beam convolution (PBC), analytical anisotropy algorithm (AAA), and Acuros XB TPS algorithms. Relative depth dose curves were calculated in heterogeneous water phantoms with layers of bone (1.8 g/cm3) and lung (0.3 g/cm3) equivalent materials for radiation fields between 1 × 1 cm2 and 10 × 10 cm2. Analysis of relative depth dose curves at the water-bone interface shows that PBC and AAA algorithms present the largest differences to MC calculations (uMC = 0.5%), with maximum differences of up to 4.3% of maximum dose. For the lung-equivalent material and 1 × 1 cm2 field, differences can be up to 24.3% for PBC, 11.5% for AAA, and 7.5% for Acuros. Results show that Acurus presents the best agreement with MC simulation data with equivalent accuracy for modeling radiotherapy dose deposition especially in regions where electronic equilibrium does not hold. For typical (nonsmall) fields used in radiotherapy, AAA and PBC can exhibit reasonable agreement with MC results even in regions of heterogeneities.
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Rational Dosage Regimen: Maintenance Dose and Loading Dose
In most cases, drugs are administered repetitively or infused continuously to maintain a steady-state concentration in the body. At a steady...
Drug Dosing: Infants and Children
Drug Dosing: Geriatric Patients