Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
Clinically recorded cortical auditory evoked potentials from paediatric cochlear implant users fitted with
Julie Kosaner1, Bram Van Dun2, Ozgur Yigit3
1Meders Speech and Hearing Clinic, Meders İşitme ve Konuşma Merkezi, Söğütlüçeşme Caddesi: No 102, Kadıköy, İstanbul 34714, Turkey.
Objectives:
This study aimed to objectively evaluate access to soft sounds (55 dB SPL) in paediatric CI users, all wearing MED-EL (Innsbruck, Austria) devices who were fitted with the objective electrically elicited stapedius reflex threshold (eSRT) fitting method, to track their cortical auditory evoked potential (CAEP) presence and latency, and to compare their CAEPs to those of normal-hearing peers.
Methods:
Forty-five unilaterally implanted, pre-lingually deafened MED-EL CI users, aged 12-48 months, underwent CAEP testing in the clinic at regular monthly intervals post switch-on. CAEPs were recorded in response to short speech tokens /m/, /g/ and /t/ presented in the free field at 55 dB SPL. Twenty children with normal hearing (NH), similarly aged, underwent CAEP testing once.
Results:
The proportion of present CAEPs increased and CAEP P1 latencies reduced significantly with post-implantation duration. CAEPs were scored based on their presence and age-appropriate P1 latency. These CAEP scores increased significantly with post-implantation duration. CAEP scores were significantly worse for the /m/ speech token compared to the other two tokens. Compared to the NH group, CAEP scores were significantly smaller for all post-implantation test intervals.
Conclusions:
This study provides clinicians with a first step towards typical ranges of CAEP presence, latency, and derived CAEP score over the first months of MED-EL CI use. CAEPs within these typical ranges could validate intervention whereas less than optimum CAEPs could prompt clinicians to seek solutions in a timely manner. CAEPs could clinically validate whether a CI provides adequate access to soft sounds. This approach could form an alternative to behavioural soft sound access verification.
Related Concept Videos
Electric Potential and Potential Difference
When a test charge moves from the initial to the final position, the electric potential difference between those positions is defined as the ratio of the change in the potential energy to the charge on the...
Finding Electric Potential From Electric Field
Calculations of Electric Potential I
The ring is divided into infinitesimal small arcs such that point M is equidistant from all the arcs. Here, the cylindrical coordinate system is used to calculate the electric potential at point M. A general element of the arc between angles θ and θ + dθ is of the length Rdθ and has a charge of...
Calculations of Electric Potential II
Consider a...
Determining Electric Field From Electric Potential
In general, regardless of whether the electric field is uniform, it points in the direction of decreasing potential because the force on a positive...
Electric Potential Energy
The electrostatic or Coulomb...