Related Experiment Video
Updated: Feb 12, 2026

Adaptation of a Haptic Robot in a 3T fMRI
Published on: October 4, 2011
Comparison of Two Toric IOLs with Different Haptic Design: Optical Quality after 1 Year
Kata Miháltz1,2, Michael Lasta1,2, Michael Burgmüller1,2
1Department of Ophthalmology, Hospital Hietzing, Vienna, Austria.
Background:
The purpose of this prospective, randomised study was to interocularly compare the visual performance after implantation of two different toric IOLs with different haptic design.
Methods:
59 subjects with corneal astigmatism greater than 1.25 diopter (D) were implanted with an AT TORBI 709M IOL (Carl Zeiss Meditec AG) in one eye and with a Tecnis toric aspheric IOL (Abbot Medical Optics) in the other eye. Observation procedure was performed 12 months postoperatively. Main outcome measures included uncorrected distance visual acuity (UDVA), manifest refraction, IOL rotation, and IOL position.
Results:
Mean UCDVA was 0.04 ± 0.14 logMAR for AT TORBI eyes and 0.06 ± 0.15 logMAR for Tecnis eyes (p = 0.3). The postoperative spherical equivalent values were significantly lower in the AT TORBI group. Mean toric IOL axis rotation was 3.0 ± 2.26 degrees for AT TORBI eyes and 3.27 ± 2.37 for Tecnis eyes (p = 0.5). The mean vertical IOL tilt and vertical decentration values measured with the Visante OCT were significantly larger in the AT TORBI group (p < 0.05).
Conclusions:
Both the Tecnis and the AT TORBI toric IOLs successfully reduced ocular astigmatism. Emmetropia could be better achieved with the AT TORBI IOL, whereas the Tecnis showed better positional stability. This trial is registered with ICMJE NCT03371576.
More Related Videos
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
13:07Convergent Polishing: A Simple, Rapid, Full Aperture Polishing Process of High Quality Optical Flats & Spheres
Published on: December 1, 2014
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Group Design
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Quality Assurance
Quality of Water
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...