Exradin W1 plastic scintillation detector for in vivo skin dosimetry in passive scattering proton therapy
Fahed Alsanea1, Landon Wootton2, Narayan Sahoo1
1Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX 77030, USA; The University of Texas Graduate School of Biomedical Sciences at Houston, Houston, TX 77030, USA.
Abstract:
In vivo skin dosimetry is desirable in passive scattering proton therapy because of the possibility of high entrance dose with a small number of fields. However, suitable detectors are needed to determine skin dose in proton therapy. Plastic scintillation detectors (PSDs) are particularly well suited for applications in proton therapy because of their water equivalence, small size, and ease of use. We investigated the utility of the Exradin W1, a commercially available PSD, for in vivo skin dosimetry during passive scattering proton therapy. We evaluated the accuracy of the Exradin W1 in six patients undergoing proton therapy for prostate cancer, as part of an Institutional Review Board-approved protocol. Over 22 weeks, we compared in vivo PSD measurements with in-phantom ionization chamber measurements and doses from the treatment planning system, resulting in 96 in vivo measurements. Temperature and ionization quenching correction factors were applied on the basis of the dose response of the PSD in a phantom. The calibrated PSD exhibited an average 7.8% under-response (±1% standard deviation) owing to ionization quenching. We observed 4% under-response at 37 °C relative to the calibration-temperature response. After temperature and quenching corrections were applied, the overall PSD dose response was within ±1% of the expected dose for all patients. The dose differences between the PSD and ionization chamber measurements for all treatment fields were within ±2% (standard deviation 0.67%). The PSD was highly accurate for in vivo skin dosimetry in passively scattered proton beams and could be useful in verifying proton therapy delivery.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Related Concept Videos
Plasticity
Plasticizers
Plasticizers function by using surface-active agents to create repulsive electrostatic forces between cement particles. This dispersion enhances the concrete's...
Passive Filters
Low-Pass Filters
Low-pass filters are designed to transmit signals with frequencies lower than the cutoff frequency, ωc, and attenuate those above it. The cutoff...
Gene Therapy
Plastic Behavior
Plastic Deformations